  Comparison of prospective head motion correction with NMR field probes and an optical tracking system

Eschelbach, M., Aghaeifar, A., Bause, J., Handwerker, J., Anders, J., Engel, E.-M., Thielscher, A., & Scheffler, K. (2019). Comparison of prospective head motion correction with NMR field probes and an optical tracking system. Magnetic Resonance in Medicine, 81(1), 719-729. doi:10.1002/mrm.27343.

Item is

基本情報

表示: 非表示:
アイテムのパーマリンク: https://hdl.handle.net/21.11116/0000-0001-EA81-5 版のパーマリンク: https://hdl.handle.net/21.11116/0000-0002-930C-B
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Eschelbach, M1, 2, 著者           
Aghaeifar, A1, 2, 著者           
Bause, J1, 2, 著者           
Handwerker, J, 著者
Anders, J, 著者
Engel, E-M, 著者
Thielscher, A1, 2, 著者           
Scheffler, K1, 2, 著者           
所属:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              
2Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497794              

内容説明

表示:
非表示:
キーワード: -
 要旨: Purpose
The aim of this study was to compare prospective head motion correction and motion tracking abilities of two tracking systems: Active NMR field probes and a Moiré phase tracking camera system using an optical marker.
Methods
Both tracking systems were used simultaneously on human subjects. The prospective head motion correction was compared in an MP2RAGE and a gradient echo sequence. In addition, the motion tracking trajectories for three subjects were compared against each other and their correlation and deviations were analyzed.
Results
With both tracking systems motion artifacts were visibly reduced. The precision of the field probe system was on the order of 50 µm for translations and 0.03° for rotations while the camera's was approximately 5 µm and 0.007°. The comparison of the measured trajectories showed close correlation and an average absolute deviation below 500 µm and 0.5°.
Conclusion
This study presents the first in vivo comparison between NMR field probes and Moiré phase tracking. For the gradient echo images, the field probes had a similar motion correction performance as the optical tracking system. For the MP2RAGE measurement, however, the camera yielded better results. Still, both tracking systems substantially decreased image artifacts in the presence of subject motion. Thus, the motion tracking modality should be chosen according to the specific requirements of the experiment while considering the desired image resolution, refresh rate, and head coil constraints.

資料詳細

表示:
非表示:
言語:
 日付: 2018-072019-01
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): DOI: 10.1002/mrm.27343
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Magnetic Resonance in Medicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: New York : Wiley-Liss
ページ: - 巻号: 81 (1) 通巻号: - 開始・終了ページ: 719 - 729 識別子(ISBN, ISSN, DOIなど): ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149