  Multicenter study of subjective acceptance during magnetic resonance imaging at 7 and 9.4 T

Rauschenberg, J., Nagel, A. M., Ladd, S. C., Theysohn, J. M., Ladd, M. E., Möller, H. E., et al. (2014). Multicenter study of subjective acceptance during magnetic resonance imaging at 7 and 9.4 T. Investigative Radiology, 49(5), 249-259. doi:10.1097/RLI.0000000000000035.

Item is

Basisdaten

einblenden: ausblenden:
Genre: Zeitschriftenartikel

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Rauschenberg, Jaane1, Autor
Nagel, Arnim M1, Autor
Ladd, Susanne C2, 3, Autor
Theysohn, Jens M2, 3, Autor
Ladd, Mark E2, 3, Autor
Möller, Harald E.4, Autor           
Trampel, Robert5, Autor           
Turner, Robert5, Autor           
Pohmann, R6, Autor           
Scheffler, K6, Autor           
Brechmann, Andre7, Autor
Stadler, Jörg7, Autor
Felder, Jörg8, Autor
Shah, N Jon8, Autor
Semmler, Wolfhard1, Autor
Affiliations:
1Department of Medical Physics in Radiology, German Cancer Research Center, Heidelberg, Germany, ou_persistent22              
2Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Germany, ou_persistent22              
3epartment of Diagnostic and Interventional Radiology and Neuroradiology, University Duisburg-Essen, University Hospital, Essen, Germany, ou_persistent22              
4Methods and Development Unit Nuclear Magnetic Resonance, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634558              
5Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634550              
6Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              
7pecial Lab Non-Invasive Brain Imaging, Leibniz Institute for Neurobiology, Magdeburg, Germany, ou_persistent22              
8Research Center Julich, Institute of Neuroscience and Medicine, Julich, Germany, ou_persistent22              

Inhalt

einblenden:
ausblenden:
Schlagwörter: MR safety; Sensory side effects; Discomfort; High field; Questionnaire
 Zusammenfassung: Objectives: The aims of this study were to investigate the subjective discomfort and sensory side effects during ultrahigh field (UHF) magnetic resonance imaging (MRI) examinations in a large-scale study and to evaluate differences between magnetic resonance (MR) sites.

Materials and Methods: Four MR sites with a 7-T MR system and 2 MR sites with a 9.4-T MR system participated in this multicenter study with a total number of 3457 completed questionnaires on causes of discomfort and sensations during the examination. For a pooled retrospective analysis of the results from the partially different questionnaires, all data were adapted to an answer option with a 4-point scale (0 = no discomfort/side effect, 3 = very unpleasant/very strong sensation). To differentiate effects evoked by the low-frequency time-varying magnetic fields due to movement through the static magnetic field, most questionnaires separated the manifestation of sensory side effects during movement on the patient table from manifestation while lying still in the isocenter.

Results: In general, a high acceptance of UHF examinations was found, where in 82% of the completed questionnaires, the subjects stated the examination to be at least tolerable. Although in 7.6% of the questionnaires, subjects felt discomfort during the examination, only 0.9% of the image acquisitions had to be terminated prematurely. No adverse events occurred in any of the examinations. Only 1% of the subjects were unwilling to undergo further UHF MRI examinations. Examination duration was the most complained cause of discomfort, followed by acoustic noise and lying still. All magnetic-field-related sensations were more pronounced when moving the patient table versus the isocenter position (19%/2% of the subjects felt unpleasant vertigo during the moving/stationary state). In general, vertigo was the most often stated sensory side effect and was more pronounced at 9.4 T compared with 7 T. However, the results varied substantially among the different sites.

Conclusions: The high levels of subjective acceptance found in this study lead to the conclusion that UHF MRI would be tolerated as a diagnostic tool in clinical practice. For more consistent data ascertainment, we propose a standardized questionnaire for subjective perception monitoring.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2014-05
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1097/RLI.0000000000000035
PMID: 24637589
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Investigative Radiology
  Andere : Invest. Radiol.
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: -
Seiten: - Band / Heft: 49 (5) Artikelnummer: - Start- / Endseite: 249 - 259 Identifikator: ISSN: 0020-9996
CoNE: https://pure.mpg.de/cone/journals/resource/954925408761