日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細

登録内容を編集ファイル形式で保存
 
 
ダウンロード電子メール
  Development of high‐resolution 3D MR fingerprinting for detection and characterization of epileptic lesions

Ma, D., Jones, S., Deshmane, A., Sakaie, K., Pierre, E., Larvie, M., McGivney, D., Blümcke, I., Krishnan, B., Lowe, M., Gulani, V., Najm, I., Griswold, M., & Wang, Z. (2019). Development of high‐resolution 3D MR fingerprinting for detection and characterization of epileptic lesions. Journal of Magnetic Resonance Imaging, 49(5), 1333-1346. doi:10.1002/jmri.26319.

Item is

基本情報

表示: 非表示:
アイテムのパーマリンク: https://hdl.handle.net/21.11116/0000-0002-B728-3 版のパーマリンク: https://hdl.handle.net/21.11116/0000-0003-5D43-A
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Ma, D, 著者
Jones, SE, 著者
Deshmane, A1, 2, 著者           
Sakaie, K, 著者
Pierre, EY, 著者
Larvie, M, 著者
McGivney, D, 著者
Blümcke, I, 著者
Krishnan, B, 著者
Lowe, M, 著者
Gulani, V, 著者
Najm, I, 著者
Griswold, MA, 著者
Wang, ZI, 著者
所属:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              
2Max Planck Institute for Biological Cybernetics, Max Planck Society, Spemannstrasse 38, 72076 Tübingen, DE, ou_1497794              

内容説明

表示:
非表示:
キーワード: -
 要旨: Background
Conventional MRI can be limited in detecting subtle epileptic lesions or identifying active/epileptic lesions among widespread, multifocal lesions.
Purpose
We developed a high‐resolution 3D MR fingerprinting (MRF) protocol to simultaneously provide quantitative T1, T2, proton density, and tissue fraction maps for detection and characterization of epileptic lesions.
Study type
Prospective.
Population
National Institute of Standards and Technology (NIST) / International Society for Magnetic Resonance in Medicine (ISMRM) phantom, five healthy volunteers and 15 patients with medically intractable epilepsy undergoing presurgical evaluation with noninvasive or invasive electroclinical data.
Field Strength/Sequence
3D MRF scans and routine clinical epilepsy MR protocols were acquired at 3 T.
Assessment
The accuracy of the T1 and T2 values were first evaluated using the NIST/ISMRM phantom. The repeatability was then estimated with both phantom and volunteers based on the coefficient of variance (CV). For epilepsy patients, all the maps were qualitatively reviewed for lesion detection by three independent reviewers (S.E.J., M.L., I.N.) blinded to clinical data. Region of interest (ROI) analysis was performed on T1 and T2 maps to quantify the multiparametric signal differences between lesion and normal tissues. Findings from qualitative review and quantitative ROI analysis were compared with patients' electroclinical data to assess concordance.
Statistical Tests
Phantom results were compared using R‐squared, and patient results were compared using linear regression models.
Results
The phantom study showed high accuracy with the standard values, with an R2 of 0.99. The volunteer study showed high repeatability, with an average CV of 4.3% for T1 and T2 in various tissue regions. For the 15 patients, MRF showed additional findings in four patients, with the remaining 11 patients showing findings consistent with conventional MRI. The additional MRF findings were highly concordant with patients' electroclinical presentation.
Data Conclusion
The 3D MRF protocol showed potential to identify otherwise inconspicuous epileptogenic lesions from the patients with negative conventional MRI diagnosis, as well as to correlate with different levels of epileptogenicity when widespread lesions were present.

資料詳細

表示:
非表示:
言語:
 日付: 2018-122019-05
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): DOI: 10.1002/jmri.26319
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Journal of Magnetic Resonance Imaging
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: Chicago, IL : Society for Magnetic Resonance Imaging
ページ: - 巻号: 49 (5) 通巻号: - 開始・終了ページ: 1333 - 1346 識別子(ISBN, ISSN, DOIなど): ISSN: 1053-1807
CoNE: https://pure.mpg.de/cone/journals/resource/954925594512