  Design of a shim coil array matched to the human brain anatomy

Jia, F., Elshatlawy, H., Aghaeifar, A., Chu, Y., Hsu, Y., Littin, S., Kroboth, S., Yu, H., Amrein, P., Gao, X., Yang, W., LeVan, P., Scheffler, K., & Zaitsev, M. (2020). Design of a shim coil array matched to the human brain anatomy. Magnetic Resonance in Medicine, 83(4), 1442-1457. doi:10.1002/mrm.28016.

Item is

基本情報

表示: 非表示:
アイテムのパーマリンク: https://hdl.handle.net/21.11116/0000-0004-C6D8-9 版のパーマリンク: https://hdl.handle.net/21.11116/0000-0005-72BC-7
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Jia, F, 著者
Elshatlawy, H, 著者
Aghaeifar, A1, 2, 著者           
Chu, YH, 著者
Hsu, YC, 著者
Littin, S, 著者
Kroboth, S, 著者
Yu, H, 著者
Amrein, P, 著者
Gao, X, 著者
Yang, W, 著者
LeVan, P, 著者
Scheffler, K1, 2, 著者           
Zaitsev, M, 著者
所属:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              
2Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497794              

内容説明

表示:
非表示:
キーワード: -
 要旨: Purpose

The purpose of this study is to introduce a novel design method of a shim coil array specifically optimized for whole brain shimming and to compare the performance of the resulting coils to conventional spherical harmonic shimming.
Methods

The proposed design approach is based on the stream function method and singular value decomposition. Eighty‐four field maps from 12 volunteers measured in seven different head positions were used during the design process. The cross validation technique was applied to find an optimal number of coil elements in the array. Additional 42 field maps from 6 further volunteers were used for an independent validation. A bootstrapping technique was used to estimate the required population size to achieve a stable coil design.
Results

Shimming using 12 and 24 coil elements outperforms fourth‐ and fifth‐order spherical harmonic shimming for all measured field maps, respectively. Coil elements show novel coil layouts compared to the conventional spherical harmonic coils and existing multi‐coils. Both leave‐one‐out and independent validation demonstrate the generalization ability of the designed arrays. The bootstrapping analysis predicts that field maps from approximately 140 subjects need to be acquired to arrive at a stable design.
Conclusions

The results demonstrate the validity of the proposed method to design a shim coil array matched to the human brain anatomy, which naturally satisfies the laws of electrodynamics. The design method may also be applied to develop new shim coil arrays matched to other human organs.

資料詳細

表示:
非表示:
言語:
 日付: 2019-092020-04
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): DOI: 10.1002/mrm.28016
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Magnetic Resonance in Medicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: New York : Wiley-Liss
ページ: - 巻号: 83 (4) 通巻号: - 開始・終了ページ: 1442 - 1457 識別子(ISBN, ISSN, DOIなど): ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149