  True real-time cardiac MRI in free breathing without ECG synchronization using a novel sequence with radial k-space sampling and balanced SSFP contrast mode.

Bauer, R. W., Radtke, I., Block, K. T., Larson, M. C., Kerl, J. M., Hammerstingl, R., et al. (2013). True real-time cardiac MRI in free breathing without ECG synchronization using a novel sequence with radial k-space sampling and balanced SSFP contrast mode. The International Journal of Cardiovascular Imaging, 29(5), 1059-1067. doi:10.1007/s10554-013-0183-0.

Item is

Dateien

einblenden: Dateien
ausblenden: Dateien
:
1833516.pdf (Verlagsversion), 435KB
 
Datei-Permalink:
-
Name:
1833516.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Eingeschränkt (UNKNOWN id 303; )
MIME-Typ / Prüfsumme:
application/pdf
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:
ausblenden:
Beschreibung:
-
OA-Status:

Urheber

einblenden:
ausblenden:
 Urheber:
Bauer, R. W., Autor
Radtke, I., Autor
Block, K. T., Autor
Larson, M. C., Autor
Kerl, J. M., Autor
Hammerstingl, R., Autor
Graf, T. G., Autor
Vogl, T. J., Autor
Zhang, S.1, Autor           
Affiliations:
1Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society, ou_578634              

Inhalt

einblenden:
ausblenden:
Schlagwörter: Real-time MRI; Cardiac MRI; Radial sampling; Cardiac function
 Zusammenfassung: We investigated a novel sequence with radial k-space sampling, gridding and sliding window reconstruction with bSSFP contrast that allows for true real-time functional cardiac evaluation independent from respiration and ECG triggering. 12 healthy volunteers underwent 1.5 T cardiac MRI. Single-shot short axis views were acquired with a) standard retrospectively ECG-gated segmented breath-hold (bh) bSSFP and with the real-time radial bSSFP sequence with a nominal temporal resolution of b) 16 fps (frames per second) and c) 40 fps. Radial bSSFP were acquired during free breathing without ECG synchronization. Left ventricular functional parameters (EDV, ESV, SV and EF) were compared and quality of wall motion depiction was assessed. Contrast-to-noise-ratio (CNR) of myocardium/blood pool in the left ventricle was calculated. EF showed excellent correlation (Bland-Altman r = 0.99; Lin rho = 0.91) between bh-bSSFP (65 %) and 40 fps radial (64 %) and moderate correlation (r = 0.84, rho = 0.20) with 16 fps radial bSSFP (56 %). While EDV was in good agreement for all three sequences, ESV was significantly overestimated with 16 fps radial bSSFP. Despite lower CNR, image quality for wall motion assessment was rated significantly better for 40 fps compared to 16 fps radial bSSFP due to the faster temporal resolution. Left ventricular functional analysis with fast true real-time radial bSSFP is in good agreement with standard ECG-gated bh-bSSFP. The independency from ECG synchronization and breathing promises a robust method for patients with impaired cardiopulmonary status in whom breath-hold and good quality ECG cannot be achieved.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2013-01-192013-06
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1007/s10554-013-0183-0
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: The International Journal of Cardiovascular Imaging
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: -
Seiten: - Band / Heft: 29 (5) Artikelnummer: - Start- / Endseite: 1059 - 1067 Identifikator: -