  Localized B0 shimming based on 23Na MRI at 7 T

Gast, L., Henning, A., Hensel, B., Uder, M., & Nagel, A. (2019). Localized B0 shimming based on 23Na MRI at 7 T. Magnetic Resonance in Medicine, Epub ahead. doi:10.1002/mrm.28011.

Item is

基本情報

表示: 非表示:
アイテムのパーマリンク: https://hdl.handle.net/21.11116/0000-0004-D27F-1 版のパーマリンク: https://hdl.handle.net/21.11116/0000-0004-D280-D
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Gast, LV, 著者
Henning, A1, 2, 著者           
Hensel, B, 著者
Uder, M, 著者
Nagel, AM, 著者
所属:
1Research Group MR Spectroscopy and Ultra-High Field Methodology, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_2528692              
2Max Planck Institute for Biological Cybernetics, Max Planck Society, Spemannstrasse 38, 72076 Tübingen, DE, ou_1497794              

内容説明

表示:
非表示:
キーワード: -
 要旨:
Purpose

To validate the feasibility of localized B0 shimming based on B0 maps acquired with sodium (23Na) MRI.
Methods

A localized B0 shimming routine based on a constrained regularized algorithm in combination with 23Na MRI data acquired with a 3D density‐adapted radial readout scheme was implemented on a 7T MR system. Measurements were performed using a dual‐tuned 23Na/1H head coil. The quality of B0 maps reconstructed from 23Na images and the resulting shim values was examined depending on the acquisition duration between 10 minutes and 15 seconds to examine clinical applicability. The B0 shimming based on 23Na B0 maps was performed both for phantom and human head of 6 healthy volunteers, and the resulting B0 homogeneity was compared with the vendor‐provided 1H MRI–based gradient‐echo brain shimming routine.
Results

The proposed 23Na MRI–based shimming routine showed a reduction in B0 variation comparable to the vendor‐provided shim both in phantom and in vivo measurements. Within the examined multicompartment phantom, the B0 variations could be reduced by up to 77% using the 23Na MRI–based shim. In human head, B0 variations were reduced by approximately 50% using an acquisition time of 15 seconds for the 23Na B0 maps and only 1 iteration of B0 shimming.
Conclusion

The 23Na MRI–based localized B0 shimming is possible at 7 T within clinically acceptable acquisition durations (< 1 minute). It was shown that using the proposed 23Na MRI–based shimming approach, the 23Na image quality at ultrahigh field strength can be strongly improved.

資料詳細

表示:
非表示:
言語:
 日付: 2019-10
 出版の状態: オンラインで出版済み
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): DOI: 10.1002/mrm.28011
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Magnetic Resonance in Medicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: New York : Wiley-Liss
ページ: - 巻号: Epub ahead 通巻号: - 開始・終了ページ: - 識別子(ISBN, ISSN, DOIなど): ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149