  A simple low-SAR technique for chemical-shift selection with high-field spin-echo imaging

Ivanov, D., Schäfer, A., Streicher, M., Heidemann, R. M., Trampel, R., & Turner, R. (2010). A simple low-SAR technique for chemical-shift selection with high-field spin-echo imaging. Magnetic Resonance in Medicine, 64(2), 319-326. doi:10.1002/mrm.22518.

Item is

Dateien

einblenden: Dateien
ausblenden: Dateien
:
A Simple Low-SAR Technique for Chemical-Shift Selection with High-Field Spin-Echo Imaging-Ivanov-Turner-2010.pdf (beliebiger Volltext), 514KB
Name:
A Simple Low-SAR Technique for Chemical-Shift Selection with High-Field Spin-Echo Imaging-Ivanov-Turner-2010.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Öffentlich
MIME-Typ / Prüfsumme:
application/pdf / [MD5]
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Ivanov, Dimo1, Autor           
Schäfer, Andreas1, Autor           
Streicher, Markus1, Autor           
Heidemann, Robin M.1, Autor           
Trampel, Robert1, Autor           
Turner, Robert1, Autor           
Affiliations:
1Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634550              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 Zusammenfassung: We have discovered a simple and highly robust method for removal of chemical shift artifact in spin-echo MR images, which simultaneously decreases the radiofrequency power deposition (specific absorption rate). The method is demonstrated in spin-echo echo-planar imaging brain images acquired at 7 T, with complete suppression of scalp fat signal. When excitation and refocusing pulses are sufficiently different in duration, and thus also different in the amplitude of their slice-select gradients, a spatial mismatch is produced between the fat slices excited and refocused, with no overlap. Because no additional radiofrequency pulse is used to suppress fat, the specific absorption rate is significantly reduced compared with conventional approaches. This enables greater volume coverage per unit time, well suited for functional and diffusion studies using spin-echo echo-planar imaging. Moreover, the method can be generally applied to any sequence involving slice-selective excitation and at least one slice-selective refocusing pulse at high magnetic field strengths. The method is more efficient than gradient reversal methods and more robust against inhomogeneities of the static (polarizing) field (B0).

Details

einblenden:
ausblenden:
Sprache(n):
 Datum: 2010
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: eDoc: 512107
Anderer: P11704
DOI: 10.1002/mrm.22518
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Magnetic Resonance in Medicine
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: New York : Wiley-Liss
Seiten: - Band / Heft: 64 (2) Artikelnummer: - Start- / Endseite: 319 - 326 Identifikator: ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149