  Setting up and operating a cryo-EM laboratory

Mills, D. J. (2021). Setting up and operating a cryo-EM laboratory. Quarterly Reviews of Biophysics, 54: e2. doi:10.1017/S003358352000013X.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Mills, Deryck J.1, Author           
Affiliations:
1Department of Structural Biology, Max Planck Institute of Biophysics, Max Planck Society, ou_2068291              

Content

show
hide
Free keywords: Cryo-electron microscopy; cryo-EM facility design; cryo-electron tomography; sample preparation; single-particle cryo-EM
 Abstract: Cryo-electron microscopy (cryo-EM) has become the technique of choice for structural biology of macromolecular assemblies, after the ‘resolution revolution’ that has occurred in this field since 2012. With a suitable instrument, an appropriate electron detector and, last but not least, a cooperative sample it is now possible to collect images from which macromolecular structures can be determined to better than 2 Å resolution, where reliable atomic models can be built. By electron tomography and sub-tomogram averaging of cryo-samples, it is also possible to reconstruct subcellular structures to sub-nanometre resolution. This review describes the infrastructure that is needed to achieve this goal. Ideally, a cryo-EM lab will have a dedicated 300 kV electron microscope for data recording and a 200 kV instrument for screening cryo-samples, both with direct electron detectors, and at least one 120 kV EM for negative-stain screening at room temperature. Added to this should be ancillary equipment for specimen preparation, including a light microscope, carbon coater, plasma cleaner, glow discharge unit, a device for fast, robotic sample freezing, liquid nitrogen storage Dewars and a ready supply of clean liquid nitrogen. In practice, of course, the available budget will determine the number and types of microscopes and how elaborate the lab can be. The cryo-EM lab should be designed with adequate space for the electron microscopes and ancillary equipment, and should allow for sufficient storage space. Each electron microscope room should be connected to the image-processing computers by fibre-optic cables for the rapid transfer of large datasets. The cryo-EM lab should be overseen by a facility manager whose responsibilities include the day-to-day tasks to ensure that all microscopes are operating perfectly, organising service and repairs to minimise downtime, and controlling the budget. Large facilities will require additional support staff who help to oversee the operation of the facility and instruct new users.

Details

show
hide
Language(s): eng - English
 Dates: 2020-11-272020-08-212020-11-302021-01-08
 Publication Status: Published online
 Pages: 16
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1017/S003358352000013X
PMID: 33413714
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Quarterly Reviews of Biophysics
  Other : Q. Rev. Biophys.
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: London : Cambridge University Press.
Pages: - Volume / Issue: 54 Sequence Number: e2 Start / End Page: - Identifier: ISSN: 0033-5835
CoNE: https://pure.mpg.de/cone/journals/resource/954925334697