English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
  Measuring three-dimensional tibiofemoral kinematics using dual-slice real-time magnetic resonance imaging.

Lin, C. C., Zhang, S., Hsu, C. Y., Frahm, J., Lu, T. W., & Shih, T. F. (2019). Measuring three-dimensional tibiofemoral kinematics using dual-slice real-time magnetic resonance imaging. Medical Physics, 46(10), 4588-4599. doi:10.1002/mp.13761.

Item is

Files

show Files
hide Files
:
3152652.pdf (Publisher version), 3MB
 
File Permalink:
-
Name:
3152652.pdf
Description:
-
OA-Status:
Visibility:
Restricted ( Max Planck Society (every institute); )
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show

Creators

show
hide
 Creators:
Lin, C. C., Author
Zhang, S.1, Author           
Hsu, C. Y., Author
Frahm, J.1, Author           
Lu, T. W., Author
Shih, T. F., Author
Affiliations:
1Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society, ou_578634              

Content

show
hide
Free keywords: Image registration; Knee; Real-time MRI; kinematics
 Abstract: PURPOSE:

The purpose of this study is to propose and evaluate a slice-to-volume registration (SVR) method integrating an advanced dual-slice real-time magnetic resonance image (MRI) and three-dimensional (3-D) MRI volume of the tibiofemoral joint for determining their 3-D kinematics.
METHODS:

The real-time and 3-D MRI of the knee were collected from twelve healthy adults at five static flexion positions and during dynamic flexion/extension movement. The 3-D positions and orientations of the femur and tibia were obtained by registering their volumetric models constructed from the 3-D MRI to dual-slice real-time MRI using an optimization process. The proposed method was quantitatively evaluated for its performance in terms of the robustness and measurement accuracy, and compared to those of a single-slice SVR method. Its repeatability in measuring knee kinematics during flexion/extension movement was also determined.
RESULTS:

In comparison to the single-slice SVR method, the dual-slice method was significantly superior, giving a successful registration rate > 95%, a bias less than 0.5 mm in translations and 0.6° in rotations and a precision less than 0.7 mm in translations and 0.9° in rotations for determining the 3-D tibiofemoral poses. For repeatability of the dual-slice SVR in measuring tibiofemoral kinematics during dynamic flexion-extension, the means of the time-averaged standard deviations were less than 0.9° for joint angles and 0.5 mm for joint translations.
CONCLUSION:

A dual-slice SVR method in conjunction with real-time MRI has been developed and evaluated for its performance in measuring 3-D kinematics of the tibiofemoral joint in twelve young adults in terms of the accuracy, robustness and repeatability. The proposed MRI-based 3-D measurement method provides a non-invasive and ionizing radiation-free approach for 3-D kinematic measurement of the tibiofemoral joint, which will be helpful for future academic and clinical applications.

Details

show
hide
Language(s): eng - English
 Dates: 2019-08-132019-10
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1002/mp.13761
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Medical Physics
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 46 (10) Sequence Number: - Start / End Page: 4588 - 4599 Identifier: -