English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
 
 
DownloadE-Mail
  A closed-loop pump-driven wire-guided flow jet for ultrafast spectroscopy of liquid samples

Picchiotti, A., Prokhorenko, V., & Miller, R. J. D. (2015). A closed-loop pump-driven wire-guided flow jet for ultrafast spectroscopy of liquid samples. Review of Scientific Instruments, 86(9): 093105. doi:10.1063/1.4929860.

Item is

Files

show Files
hide Files
:
1.4929860.pdf (Publisher version), 4MB
Name:
1.4929860.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
2015
Copyright Info:
© American Institute of Physics. This article may be downloaded for personal use only. Any other use requires prior permission of the author and the American Institute of Physics.
License:
-

Locators

show
hide
Locator:
http://dx.doi.org/10.1063/1.4929860 (Publisher version)
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Picchiotti, Alessandra1, 2, Author           
Prokhorenko, Valentyn1, Author           
Miller, R. J. Dwayne1, 3, Author           
Affiliations:
1Miller Group, Atomically Resolved Dynamics Department, Max Planck Institute for the Structure and Dynamics of Matter, Max Planck Society, ou_1938288              
2International Max Planck Research School for Ultrafast Imaging & Structural Dynamics (IMPRS-UFAST), Max Planck Institute for the Structure and Dynamics of Matter, Max Planck Society, ou_2266714              
3Departments of Chemistry and Physics, University of Toronto, Toronto, Ontario M5S 3H6, Canada, ou_persistent22              

Content

show
hide
Free keywords: Solvents; Ultraviolet light; Flow instabilities; Liquid thin films; Femtosecond pump probe studies
 Abstract: We describe the design and provide the results of the full characterization of a closed-loop pump-driven wire-guided flow jet system. The jet has excellent optical quality with a wide range of liquids spanning from alcohol to water based solutions, including phosphate buffers used for biological samples. The thickness of the jet film varies depending on the flow rate between 90 μm and 370 μm. The liquid film is very stable, and its thickness varies only by 0.76% under optimal conditions. Measured transmitted signal reveals a long term optical stability (hours) with a RMS of 0.8%, less than the overall noise of the spectroscopy setup used in our experiments. The closed loop nature of the overall jet design has been optimized for the study of precious biological samples, in limited volumes, to remove window contributions from spectroscopic observables. This feature is particularly important for femtosecond studies in the UV range.

Details

show
hide
Language(s): eng - English
 Dates: 2015-06-112015-08-182015-09-022015-09
 Publication Status: Issued
 Pages: 6
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1063/1.4929860
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Review of Scientific Instruments
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Melville, NY : AIP Publishing
Pages: - Volume / Issue: 86 (9) Sequence Number: 093105 Start / End Page: - Identifier: ISSN: 0034-6748
CoNE: https://pure.mpg.de/cone/journals/resource/991042742033452