English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
 
 
DownloadE-Mail
  Rapid diffusion-weighted magnetic resonance imaging of the brain without susceptibility artifacts: Single-shot STEAM with radial undersampling and iterative reconstruction.

Merrem, A., Hofer, S., Voit, D., Merboldt, K. D., Klosowski, J., Untenberger, M., et al. (2017). Rapid diffusion-weighted magnetic resonance imaging of the brain without susceptibility artifacts: Single-shot STEAM with radial undersampling and iterative reconstruction. Investigative Radiology, 52(7), 428-433. doi:10.1097/RLI.0000000000000357.

Item is

Files

show Files
hide Files
:
2434072.pdf (Publisher version), 592KB
Name:
2434072.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show

Creators

show
hide
 Creators:
Merrem, A.1, Author           
Hofer, S.1, Author           
Voit, D.1, Author           
Merboldt, K. D.1, Author           
Klosowski, J.1, Author           
Untenberger, M.1, Author           
Fleischhammer, J., Author
Frahm, J.1, Author           
Affiliations:
1Biomedical NMR Research GmbH, MPI for Biophysical Chemistry, Max Planck Society, ou_578634              

Content

show
hide
Free keywords: Diffusion-weighted MRI; Radial sampling; Iterative reconstruction; Gauss-Newton method
 Abstract: OBJECTIVE: The aim of this study was to develop a rapid diffusion-weighted (DW) magnetic resonance imaging (MRI) technique for whole-brain studies without susceptibility artifacts and measuring times below 3 minutes. MATERIALS AND METHODS: The proposed method combines a DW spin-echo module with a single-shot stimulated echo acquisition mode MRI sequence. Previous deficiencies in image quality due to limited signal-to-noise ratio are compensated for (1) by radial undersampling to enhance the flip angle and thus the signal strength of stimulated echoes; (2) by defining the image reconstruction as a nonlinear inverse problem, which is solved by the iteratively regularized Gauss-Newton method; and (3) by denoising with use of a modified nonlocal means filter. The method was implemented on a 3 T MRI system (64-channel head coil, 80 mT · m gradients) and evaluated for 10 healthy subjects and 2 patients with an ischemic lesion and epidermoid cyst, respectively. RESULTS: High-quality mean DW images of the entire brain were obtained by acquiring 1 non-DW image and 6 DW images with different diffusion directions at b = 1000 s · mm. The achievable resolution for a total measuring time of 84 seconds was 1.5 mm in plane with a section thickness of 4 mm (55 sections). A measuring time of 168 seconds allowed for an in-plane resolution of 1.25 mm and a section thickness of 3 mm (54 sections). Apparent diffusion coefficient values were in agreement with literature data. CONCLUSIONS: The proposed method for DW MRI offers immunity against susceptibility problems, high spatial resolution, adequate signal-to-noise ratio and clinically feasible scan times of less than 3 minutes for whole-brain studies. More extended clinical trials require accelerated computation and online reconstruction.

Details

show
hide
Language(s): eng - English
 Dates: 2017-02-012017-07
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1097/RLI.0000000000000357
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Investigative Radiology
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 52 (7) Sequence Number: - Start / End Page: 428 - 433 Identifier: -