  Poly(vinyl alcohol)-coated microfluidic devices for high- performance microchip electrophoresis

Belder, D., Deege, A., Kohler, F., & Ludwig, M. (2002). Poly(vinyl alcohol)-coated microfluidic devices for high- performance microchip electrophoresis. Electrophoresis, 23(20), 3567-3573. doi:10.1002/1522-2683(200210)23:20<3567:AID-ELPS3567>3.0.CO;2-3.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Belder, Detlev1, 著者           
Deege, Alfred1, 著者           
Kohler, Frank1, 著者           
Ludwig, Martin1, 著者           
所属:
1Research Group Belder, Max-Planck-Institut für Kohlenforschung, Max Planck Society, ou_1445595              

内容説明

表示:
非表示:
キーワード: Coating; Electroosmosis; Electrophoresis; Microchip; Poly(vinyl alcohol)
 要旨: The channels of microfluidic glass chips have been coated with poly(vinyl alcohol) (PVA). Applied for microchip electrophoresis, the coated devices exhibited a suppressed electroosmotic flow and improved separation performance. The superior performance of PVA-coated channels could be demonstrated by electrophoretic separations of labeled amines and by video microscopy. While a distorted sample zone is injected using uncoated channels the application of PVA-coated channels results in an improved shape of the sample zone with less band broadening. Applying PVA-coated microchips for the separation of amines labeled with Alexa Fluor 350TM even sub- second separations, utilizing a separation length of only 650 mum, could be obtained, while this was not possible using uncoated devices. By using PVA-coated devices rather than an uncoated chip a threefold increase in separation efficiencies could be observed. As the electroosmotic flow (EOF) was suppressed, the anionic compounds were detected at the anode whereas the dominant EOF in uncoated devices resulted in an effective mobility to the cathode. Besides improved separation performance another important feature of the PVA-coated channels was the suppressed adsorption of fluorescent compounds in repetitive runs which results in an improved robustness and detection sensitivity. Applying PVA-coated channels, rinsing or etching steps could be omitted while this was necessary for a reliable operation of uncoated devices.

資料詳細

表示:
非表示:
言語: eng - English
 日付: 2002-10-20
 出版の状態: 出版
 ページ: 7
 出版情報: -
 目次: -
 査読: 査読あり
 識別子(DOI, ISBNなど): DOI: 10.1002/1522-2683(200210)23:20<3567::AID-ELPS3567>3.0.CO;2-3
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Electrophoresis
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: Weinheim, Germany : Verlag Chemie,
ページ: - 巻号: 23 (20) 通巻号: - 開始・終了ページ: 3567 - 3573 識別子(ISBN, ISSN, DOIなど): ISSN: 0173-0835
CoNE: https://pure.mpg.de/cone/journals/resource/954925486765