  Novel silver-tubing method for quantitative introduction of water into high-temperature conversion systems for stable hydrogen and oxygen isotopic measurements

Qi, H. P., Groning, M., Coplen, T. B., Buck, B., Mroczkowski, S. J., Brand, W. A., et al. (2010). Novel silver-tubing method for quantitative introduction of water into high-temperature conversion systems for stable hydrogen and oxygen isotopic measurements. Rapid Communications in Mass Spectrometry, 24(13), 1821-1827. doi:10.1002/rcm.4559.

Item is

Basisdaten

einblenden: ausblenden:
Genre: Zeitschriftenartikel

Dateien

einblenden: Dateien
ausblenden: Dateien
:
BGC1375.pdf (Verlagsversion), 290KB
 
Datei-Permalink:
-
Name:
BGC1375.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Eingeschränkt (Max Planck Institute for Biogeochemistry, MJBK; )
MIME-Typ / Prüfsumme:
application/octet-stream
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Qi, H. P., Autor
Groning, M., Autor
Coplen, T. B., Autor
Buck, B., Autor
Mroczkowski, S. J., Autor
Brand, Willi A.1, Autor           
Geilmann, Heike1, Autor           
Gehre, M., Autor
Affiliations:
1Service Facility Stable Isotope, Dr. W. A. Brand, Max Planck Institute for Biogeochemistry, Max Planck Society, ou_1497772              

Inhalt

einblenden:
ausblenden:
Schlagwörter: organic-substances delta-o-18 flow pyrolysis precision nitrate
 Zusammenfassung: A new method to seal water in silver tubes for use in a TC/EA (thermal conversion/elemental analyzer) reduction unit using a semi-automated sealing apparatus can yield reproducibilities (1 standard deviation) of delta H-2 and delta O-18 measurements of 1.0 parts per thousand and 0.06 parts per thousand, respectively. These silver tubes containing reference waters may be preferred for the calibration of H- and O-bearing materials analyzed with a TC/EA reduction unit. The new sealing apparatus employs a computer-controlled stepping motor to produce silver tubes identical in length. The reproducibility of the mass of water sealed in tubes (in a range of 200-400 mu g) can be as good as 1%. Approximately 99% of the sealed silver tubes are satisfactory (leak free). Although silver tubes sealed with reference waters are robust and can be shaken or heated to 110 degrees C with no loss of integrity, they should not be frozen because the expansion during the phase transition of water to ice will break the cold seals and all the water will be lost. The tubes should be shipped in insulated containers. This new method eliminates air inclusions and isotopic fractionation of water associated with the loading of water into capsules using a syringe. The method is also more than an order of magnitude faster than preparing water samples in ordinary Ag capsules. Nevertheless, some laboratories may prefer loading water into silver capsules because expensive equipment is not needed, but users of this method are cautioned to apply the necessary corrections for evaporation, back exchange with laboratory atmospheric moisture, and blanks. Copyright (C) 2010 John Wiley & Sons, Ltd.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2010
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: DOI: 10.1002/rcm.4559
ISI: ://000279375700001
Anderer: BGC1375
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Rapid Communications in Mass Spectrometry
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: New York, NY : John Wiley & Sons
Seiten: - Band / Heft: 24 (13) Artikelnummer: - Start- / Endseite: 1821 - 1827 Identifikator: CoNE: https://pure.mpg.de/cone/journals/resource/954925574961
ISSN: 0951-4198