  Internal ventilation system of MR scanners induces specific EEG artifact during simultaneous EEG-fMRI

Nierhaus, T., Gundlach, C., Goltz, D., Thiel, S., Pleger, B., & Villringer, A. (2013). Internal ventilation system of MR scanners induces specific EEG artifact during simultaneous EEG-fMRI. NeuroImage, 74, 70-76. doi:10.1016/j.neuroimage.2013.02.016.

Item is

Basisdaten

einblenden: ausblenden:
Genre: Zeitschriftenartikel

Dateien

einblenden: Dateien
ausblenden: Dateien
:
Nierhaus_2013_Internal.pdf (Verlagsversion), 2MB
 
Datei-Permalink:
-
Name:
Nierhaus_2013_Internal.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Privat
MIME-Typ / Prüfsumme:
application/pdf
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Nierhaus, Till1, 2, 3, 4, Autor           
Gundlach, Christopher1, Autor           
Goltz, Dominique1, 5, Autor           
Thiel, Sabrina1, 3, 6, Autor           
Pleger, Burkhard1, 7, Autor           
Villringer, Arno1, 2, 3, 4, 7, Autor           
Affiliations:
1Department Neurology, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634549              
2Leipzig Research Center for Civilization Diseases (LIFE), University of Leipzig, Germany, ou_persistent22              
3Berlin School of Mind and Brain, Humboldt University Berlin, Germany, ou_persistent22              
4Berlin Neuroimaging Center, Charité University Medicine Berlin, Germany, ou_persistent22              
5Department of Experimental Psychology and Methods, University of Leipzig, Germany, ou_persistent22              
6Department of Mathematics, Humboldt University Berlin, Germany, ou_persistent22              
7Clinic for Cognitive Neurology, University of Leipzig, Germany, ou_persistent22              

Inhalt

einblenden:
ausblenden:
Schlagwörter: Simultaneous EEG-fMRI; Ventilation; Gradient artifact; Ballistocardiogram; Quantification; Gamma activity; Helium pump
 Zusammenfassung: During simultaneous EEG-fMRI acquisition, the EEG signal suffers from tremendous artifacts caused by the scanner “environment”. Particularly, gradient artifacts and the ballistocardiogram have been well characterized, along with methods to eliminate them. Here, we describe another systematic artifact in the EEG signal, which is induced by the internal ventilation system of Siemens TRIO and VERIO MR scanners. A ventilation-level dependent vibration induces specific peaks in the frequency spectrum of the EEG. These frequency peaks are in the range of physiologically relevant brain rhythms (gamma frequency range), and thus interfere with their reliable acquisition. This ventilation dependent artifact was most prominent on the electrodes placed directly on the subject's head, so it is not sufficient to simply place the EEG's amplifier outside the scanner tube. Instead, the ventilator must be switched off to fully eliminate the ventilator's artificial manipulation of EEG recordings. Without the internal ventilator system being on, the temperature within the scanner tube may rise, thus requiring shorter scanning sessions or an additional external ventilation system.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2013-02-032013-02-192013-07-01
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1016/j.neuroimage.2013.02.016
PMID: 23435207
Anderer: Epub 2013
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: NeuroImage
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: Orlando, FL : Academic Press
Seiten: - Band / Heft: 74 Artikelnummer: - Start- / Endseite: 70 - 76 Identifikator: ISSN: 1053-8119
CoNE: https://pure.mpg.de/cone/journals/resource/954922650166