  Internal ventilation system of MR scanners induces specific EEG artifact during simultaneous EEG-fMRI

Nierhaus, T., Gundlach, C., Goltz, D., Thiel, S., Pleger, B., & Villringer, A. (2013). Internal ventilation system of MR scanners induces specific EEG artifact during simultaneous EEG-fMRI. NeuroImage, 74, 70-76. doi:10.1016/j.neuroimage.2013.02.016.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル
非表示: ファイル
:
Nierhaus_2013_Internal.pdf (出版社版), 2MB
 
ファイルのパーマリンク:
-
ファイル名:
Nierhaus_2013_Internal.pdf
説明:
-
OA-Status:
閲覧制限:
非公開
MIMEタイプ / チェックサム:
application/pdf
技術的なメタデータ:
著作権日付:
-
著作権情報:
-
CCライセンス:
-

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Nierhaus, Till1, 2, 3, 4, 著者           
Gundlach, Christopher1, 著者           
Goltz, Dominique1, 5, 著者           
Thiel, Sabrina1, 3, 6, 著者           
Pleger, Burkhard1, 7, 著者           
Villringer, Arno1, 2, 3, 4, 7, 著者           
所属:
1Department Neurology, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634549              
2Leipzig Research Center for Civilization Diseases (LIFE), University of Leipzig, Germany, ou_persistent22              
3Berlin School of Mind and Brain, Humboldt University Berlin, Germany, ou_persistent22              
4Berlin Neuroimaging Center, Charité University Medicine Berlin, Germany, ou_persistent22              
5Department of Experimental Psychology and Methods, University of Leipzig, Germany, ou_persistent22              
6Department of Mathematics, Humboldt University Berlin, Germany, ou_persistent22              
7Clinic for Cognitive Neurology, University of Leipzig, Germany, ou_persistent22              

内容説明

表示:
非表示:
キーワード: Simultaneous EEG-fMRI; Ventilation; Gradient artifact; Ballistocardiogram; Quantification; Gamma activity; Helium pump
 要旨: During simultaneous EEG-fMRI acquisition, the EEG signal suffers from tremendous artifacts caused by the scanner “environment”. Particularly, gradient artifacts and the ballistocardiogram have been well characterized, along with methods to eliminate them. Here, we describe another systematic artifact in the EEG signal, which is induced by the internal ventilation system of Siemens TRIO and VERIO MR scanners. A ventilation-level dependent vibration induces specific peaks in the frequency spectrum of the EEG. These frequency peaks are in the range of physiologically relevant brain rhythms (gamma frequency range), and thus interfere with their reliable acquisition. This ventilation dependent artifact was most prominent on the electrodes placed directly on the subject's head, so it is not sufficient to simply place the EEG's amplifier outside the scanner tube. Instead, the ventilator must be switched off to fully eliminate the ventilator's artificial manipulation of EEG recordings. Without the internal ventilator system being on, the temperature within the scanner tube may rise, thus requiring shorter scanning sessions or an additional external ventilation system.

資料詳細

表示:
非表示:
言語: eng - English
 日付: 2013-02-032013-02-192013-07-01
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: 査読あり
 識別子(DOI, ISBNなど): DOI: 10.1016/j.neuroimage.2013.02.016
PMID: 23435207
その他: Epub 2013
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: NeuroImage
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: Orlando, FL : Academic Press
ページ: - 巻号: 74 通巻号: - 開始・終了ページ: 70 - 76 識別子(ISBN, ISSN, DOIなど): ISSN: 1053-8119
CoNE: https://pure.mpg.de/cone/journals/resource/954922650166