English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
  Combining 3D tracking and surgical instrumentation to determine the stiffness of spinal motion segments: A validation study

Reutlinger, C., Gédet, P., Büchler, P., Kowal, J., Rudolph, T., Burger, J., et al. (2011). Combining 3D tracking and surgical instrumentation to determine the stiffness of spinal motion segments: A validation study. Medical Engineering & Physics, 33(3), 340-346. doi:10.1016/j.medengphy.2010.10.021.

Item is

Files

show Files

Locators

show
hide
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Reutlinger, C, Author
Gédet , P, Author
Büchler, P, Author
Kowal, J, Author
Rudolph, T, Author
Burger, J, Author
Scheffler, K1, Author           
Hasler, C, Author
Affiliations:
1Institute of Radiology, University of Basel Hospital, ou_persistent22              

Content

show
hide
Free keywords: -
 Abstract: The spine is a complex structure that provides motion in three directions: flexion and extension, lateral bending and axial rotation. So far, the investigation of the mechanical and kinematic behavior of the basic unit of the spine, a motion segment, is predominantly a domain of in vitro experiments on spinal loading simulators. Most existing approaches to measure spinal stiffness intraoperatively in an in vivo environment use a distractor. However, these concepts usually assume a planar loading and motion. The objective of our study was to develop and validate an apparatus, that allows to perform intraoperative in vivo measurements to determine both the applied force and the resulting motion in three dimensional space. The proposed setup combines force measurement with an instrumented distractor and motion tracking with an optoelectronic system. As the orientation of the applied force and the three dimensional motion is known, not only force–displacement, but also moment-angle relations could be determined. The validation was performed using three cadaveric lumbar ovine spines. The lateral bending stiffness of two motion segments per specimen was determined with the proposed concept and compared with the stiffness acquired on a spinal loading simulator which was considered to be gold standard. The mean values of the stiffness computed with the proposed concept were within a range of ±15 compared to data obtained with the spinal loading simulator under applied loads of less than 5 Nm.

Details

show
hide
Language(s):
 Dates: 2011-04
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.1016/j.medengphy.2010.10.021
BibTex Citekey: ReutlingerGBKRBSH2011
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Medical Engineering & Physics
  Other : Med. Eng. Phys.
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Oxford, UK : Elsevier
Pages: - Volume / Issue: 33 (3) Sequence Number: - Start / End Page: 340 - 346 Identifier: ISSN: 1350-4533
CoNE: https://pure.mpg.de/cone/journals/resource/954927612910