  A polymer-based MR-compatible guidewire: A study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA)

Mekle, R., Hofmann, E., Scheffler, K., & Bilecen, D. (2006). A polymer-based MR-compatible guidewire: A study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA). Journal of Magnetic Resonance Imaging, 23(2), 145-155. doi:10.1002/jmri.20486.

Item is

Basic

show hide
Item Permalink: http://hdl.handle.net/11858/00-001M-0000-0013-D2D3-B Version Permalink: http://hdl.handle.net/21.11116/0000-0004-9606-C
Genre: Journal Article

Files

show Files

Locators

show
hide
Description:
-

Creators

show
hide
 Creators:
Mekle, R, Author
Hofmann, E, Author
Scheffler, K1, Author              
Bilecen, D, Author
Affiliations:
1External Organizations, ou_persistent22              

Content

show
hide
Free keywords: -
 Abstract: Purpose To introduce a newly developed polymer‐based and magnetic resonance (MR)‐compatible guidewire and to explore its capabilities with respect to interventional peripheral magnetic resonance angiography (ipMRA) in a flow phantom. Materials and Methods The guidewire is based on a polyetheretherketone (PEEK) polymer core, and small iron particles are embedded in its coating. A passive device tracking technique was designed utilizing a susceptibility artifact induced by the wire in images acquired with a balanced steady‐state free precession (b‐SSFP) sequence using small flip angles. The position of the guidewire tip was determined from image intensity maxima and overlayed onto a roadmap in near real‐time. Guidewire tracking and balloon angioplasty of an artificial stenosis were attempted in two configurations of a flow phantom. Results Successful passive guidewire tracking was performed for all phantom configurations. Robustness and accuracy of the tracking technique were sufficient for phantom studies. A balloon catheter was placed into the stenosis using the guidewire under complete MR guidance, and subsequent balloon angioplasty yielded improved flow conditions. Conclusion The new guidewire is well‐suited for clinical application due to an absence of the risk of core fracture and its atraumatic flexible tip. It opens novel prospects for the realization of ipMRA in humans that need to be explored in further studies.

Details

show
hide
Language(s):
 Dates: 2006-02
 Publication Status: Published in print
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Method: -
 Identifiers: DOI: 10.1002/jmri.20486
BibTex Citekey: MekleHSB2006
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Journal of Magnetic Resonance Imaging
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Chicago, IL : Society for Magnetic Resonance Imaging
Pages: - Volume / Issue: 23 (2) Sequence Number: - Start / End Page: 145 - 155 Identifier: ISSN: 1053-1807
CoNE: https://pure.mpg.de/cone/journals/resource/954925594512