  Balanced alternating steady-state elastography

Bieri, O., Maderwald, S., Ladd, M., & Scheffler, K. (2006). Balanced alternating steady-state elastography. Magnetic Resonance in Medicine, 55(2), 233-241. doi:10.1002/mrm.20812.

Item is

Files

show Files

Locators

show
hide
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Bieri, O, Author
Maderwald , S, Author
Ladd, ME, Author
Scheffler, K1, Author           
Affiliations:
1External Organizations, ou_persistent22              

Content

show
hide
Free keywords: -
 Abstract: A conventional balanced steady-state free precession (b-SSFP) sequence scheme was modified such that the dynamic equilibrium becomes very sensitive to small cyclic displacements, generating two distinct and alternating steady states. This novel technique is proposed for the visualization of propagating transverse acoustic shear waves, as used in MR elastography (MRE) to determine the mechanical properties of materials or in vivo soft tissue. Experiments with tissue-like agarose gel phantoms and simulations demonstrate that the novel sequence offers an increase in phase sensitivity by about one order in magnitude compared to standard motion-encoding methods. In addition, the new method benefits from the very short acquisition times achieved by b-SSFP protocols.

Details

show
hide
Language(s):
 Dates: 2006-02
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.1002/mrm.20812
BibTex Citekey: BieriMLS2006
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Magnetic Resonance in Medicine
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: New York : Wiley-Liss
Pages: - Volume / Issue: 55 (2) Sequence Number: - Start / End Page: 233 - 241 Identifier: ISSN: 0740-3194