  Continuous arterial spin labeling (CASL) setup for the primate brain at 7T using a three-coil approach

Zappe, A.-C., Merkle, H., Logothetis, N., & Pfeuffer, J. (2004). Continuous arterial spin labeling (CASL) setup for the primate brain at 7T using a three-coil approach. In 5. Neurowissenschaftliche Nachwuchskonferenz Tübingen (NeNa 2004) (pp. 4).

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Zappe, A-C1, 2, Author           
Merkle, H, Author           
Logothetis, NK1, 2, Author           
Pfeuffer, J1, 2, Author           
Affiliations:
1Department Physiology of Cognitive Processes, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497798              
2Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497794              

Content

show
hide
Free keywords: -
 Abstract: Arterial spin labeling is commonly used to measure cerebral blood flow (CBF). Since CBF signal is intrinsically low, optimization of the signal-to-noise ratio (SNR) is critical. CASL
with a separate labeling coil has the advantages of increased SNR, multi-slice capability,
and absence of magnetization transfer. Here, we report the development of CASL for
use on monkeys. Increased sensitivity was achieved by utilization of a custom-made
three-coil setup at high magnetic field (7T) with its favorable longer T1 values. The
feasibility of this approach is demonstrated with a flow phantom and with initial
experiments in the monkey.
Measurements were performed on a vertical 7T/60cm system dedicated for monkeys, which was
equipped with a second 1H transmit channel. A saddle-shaped volume coil was applied for RF
transmission and surface coil was used for RF reception. For spin tagging at carotid- and
vertebrate arteries a concavely-shaped dual loop coil set was built. Its design was based on
angiographic scans and space restriction in the neck. All three RF coils were actively-decoupled
and switched with a self-built logic unit and current driver. The flow phantom consisted of a
saline-filled bottle with two flow tubes. Flow was adjustable from 20cm/s to 100cm/s.
Single-shot, multi-slice GE-EPI was acquired at 1x1x2mm3 resolution (128x64x5,
TR/TE=3500/12ms). A 2s labeling period was followed by a variable post-labeling delay prior to
the imaging module. For CASL, the labeling frequency was switched in interleaved scans. The
CASL three-coil setup was validated on the flow phantom. Maximum label efficiencies of 0.85
were reached with a labeling power of 0.3 and 2W on the phantom and in the monkey,
respectively. CBF maps were aquired in the monkey with variable labeling periods and postlabeling
delays. These first in vivo data demonstrate feasibility and potential for future functional
studies to measure CBF, BOLD and CMRO2 changes.

Details

show
hide
Language(s):
 Dates: 2004-08
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: BibTex Citekey: ZappeMLP2004
 Degree: -

Event

show
hide
Title: 5. Neurowissenschaftliche Nachwuchskonferenz Tübingen (NeNa 2004)
Place of Event: Oberjoch, Germany
Start-/End Date: -

Legal Case

show

Project information

show

Source 1

show
hide
Title: 5. Neurowissenschaftliche Nachwuchskonferenz Tübingen (NeNa 2004)
Source Genre: Proceedings
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: - Sequence Number: - Start / End Page: 4 Identifier: -