  Transient signal analysis using complementary metal oxide semiconductor capacitive chemical microsensors.

Kummer, A. M., Burg, T. P., & Hierlemann, A. (2006). Transient signal analysis using complementary metal oxide semiconductor capacitive chemical microsensors. Analytical Chemistry, 78(1), 279-290. doi:10.1021/ac051430g.

Item is

Files

show Files
hide Files
:
1851233.pdf (Publisher version), 382KB
 
File Permalink:
-
Name:
1851233.pdf
Description:
-
OA-Status:
Visibility:
Restricted (UNKNOWN id 303; )
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-
:
1851233_Supplement_1.htm (Supplementary material), 341KB
Name:
1851233_Supplement_1.htm
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
text/html / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show
hide
Locator:
http://pubs.acs.org/doi/pdf/10.1021/ac051430g (Publisher version)
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Kummer, A. M., Author
Burg, T. P.1, Author           
Hierlemann, A., Author
Affiliations:
1Research Group of Biological Micro- and Nanotechnology, MPI for biophysical chemistry, Max Planck Society, ou_578602              

Content

show
hide
Free keywords: -
 Abstract: This work explores the possibility to discriminate analytes based on their nonequilibrium signals in polymer-coated capacitive chemical microsensors. The analyte uptake in the chemically sensitive polymer layers of 3−7-μm thickness has been analyzed using a diffusion model and the dynamic sensor response data. The shapes of the response profiles have been calculated analytically. Despite the simplifications in the model, the observed transient signal profiles could be described accurately. Comparison of the measured diffusion coefficients (on the order of 10-12 m2/s) with literature values measured at similar concentration levels showed good agreement. Concentration-independent diffusion coefficients for several analyte/polymer combinations (poly(etherurethane)/all analytes; poly(epichlorohydrin)/alcohols) as well as slightly concentration-dependent diffusion coefficients (poly(epichlorohydrin)/toluene or ethyl cellulose/toluene) have been found in the investigated concentration range of tens to hundreds of pascals gas-phase partial pressure. The diffusion times of water and the first aliphatic monohydric alcohols in the polymers are strongly correlated to their molecular size. The discrimination of these substances based on dynamic sensor data of a single sensor could be demonstrated. In particular, the analysis of mixtures of analytes with similar chemical behavior (water/ethanol or methanol/ethanol) by means of analyzing the response profile of single-exposure steps or by applying a series of decreasingly long alternating target gas exposure and carrier gas exposure steps has been performed.

Details

show
hide
Language(s): eng - English
 Dates: 2005-12-022006
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1021/ac051430g
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Analytical Chemistry
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 78 (1) Sequence Number: - Start / End Page: 279 - 290 Identifier: -