  GIOVE, a shallow laboratory Ge-spectrometer with 100 μBq/kg sensitivity

Denz, T., Hakenmueller, J., Heusser, G., Hofacker, R., & Weber, M. (2013). GIOVE, a shallow laboratory Ge-spectrometer with 100 μBq/kg sensitivity. American Institute of Physics Conference Proceedings, 1549: 12. doi:10.1063/1.4818065.

Item is

Files

show Files
hide Files
:
GioveAIP_proc2013.pdf (Any fulltext), 1001KB
Name:
GioveAIP_proc2013.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show
hide
Locator:
http://dx.doi.org/10.1063/1.4818065 (Publisher version)
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Denz , Tobias1, Author
Hakenmueller, Janina1, Author           
Heusser, Gerd1, Author           
Hofacker , Reinhard1, Author
Weber, Marc1, Author           
Affiliations:
1Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society, ou_904549              

Content

show
hide
Free keywords: -
 Abstract: A new germanium gamma spectrometer called GIOVE ( G ermanium spectrometer with I nner and O uter V eto) has been set up at the underground/shallow laboratory (15 m w.e.) of MPI-K. Its double plastic scintillator veto system and neutron moderation interlayer lower the background by more than one order of magnitude compared to the other existing spectrometer at this facility. The integral (40-2700 keV) background rate of about 290 counts (day kg)−1 is just a factor 4 to 8 above that of the GeMPI spectrometers operated at LNGS (3800 m w.e.) and thus proves that even under shallow overburden sub mBq/kg sensitivities are achievable. Extended material screening and neutron attenuation studies preceded the final design of the spectrometer. The technical realization of the spectrometer is described in detail with special emphasis on the inner veto system. For its optimisation a simulation model was developed for light collection on small low activity PMT’s under various geometrical conditions. Radon suppression is accomplished by employing a gas tight sample container and a nitrogen flushed glove-box system with an airlock. The active volume of the crystal was modelled by absorption scanning measurements and Monte Carlo simulations. The complete shield is implemented in a Geant4 based simulation framework.

Details

show
hide
Language(s): eng - English
 Dates: 2013-08-08
 Publication Status: Published online
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.1063/1.4818065
 Degree: -

Event

show
hide
Title: LOW RADIOACTIVITY TECHNIQUES 2013 (LRT 2013)
Place of Event: Laboratori del Gran Sasso, Assergi, Italy
Start-/End Date: 2013-04-10 - 2013-04-12

Legal Case

show

Project information

show

Source 1

show
hide
Title: American Institute of Physics Conference Proceedings
  Other : AIP Conf. Proc.
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: New York : American Institute of Physics
Pages: - Volume / Issue: 1549 Sequence Number: 12 Start / End Page: - Identifier: ISSN: 0094-243X
CoNE: https://pure.mpg.de/cone/journals/resource/954928528968