  First-order perturbative calculation of the frequency-shifts caused by static cylindrically-symmetric electric and magnetic imperfections of a Penning trap

Ketter, J., Eronen, T., Höcker, M., Streubel, S., & Blaum, K. (2014). First-order perturbative calculation of the frequency-shifts caused by static cylindrically-symmetric electric and magnetic imperfections of a Penning trap. International Journal of Mass Spectrometry, 358, 1-16. doi:10.1016/j.ijms.2013.10.005.

Item is

Files

show Files
hide Files
:
1305.4861v2 (Preprint), 423KB
Name:
1305.4861v2
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show
hide
Locator:
http://dx.doi.org/10.1016/j.ijms.2013.10.005 (Publisher version)
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Ketter, Jochen1, Author           
Eronen, Tommi1, Author           
Höcker, Martin1, Author           
Streubel, Sebastian1, Author           
Blaum, Klaus1, Author           
Affiliations:
1Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              

Content

show
hide
Free keywords: -
 MPINP: Präzisionsexperimente - Abteilung Blaum
 Abstract: The ideal Penning trap consists of a uniform magnetic field and an electrostatic quadrupole potential. Cylindrically-symmetric deviations thereof are parametrized by the coefficients Bη and Cη, respectively. Relativistic mass-increase aside, the three characteristic eigenfrequencies of a charged particle stored in an ideal Penning trap are independent of the three motional amplitudes. This threefold harmonicity is a highly-coveted virtue for precision experiments that rely on the measurement of at least one eigenfrequency in order to determine fundamental properties of the stored particle, such as its mass. However, higher-order contributions to the ideal fields result in amplitude-dependent frequency-shifts. In turn, these frequency-shifts need to be understood for estimating systematic experimental errors, and eventually for correcting them by means of calibrating the imperfections. The problem of calculating the frequency-shifts caused by small imperfections of a near-ideal trap yields nicely to perturbation theory, producing analytic formulas that are easy to evaluate for the relevant parameters of an experiment. In particular, the frequency-shifts can be understood on physical rather than purely mathematical grounds by considering which terms actually drive them. Based on identifying these terms, we derive general formulas for the first-order frequency-shifts caused by any perturbation parameter Bη or Cη.

Details

show
hide
Language(s): eng - English
 Dates: 2014-01-15
 Publication Status: Issued
 Pages: 16
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1016/j.ijms.2013.10.005
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: International Journal of Mass Spectrometry
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Amsterdam : Elsevier
Pages: - Volume / Issue: 358 Sequence Number: - Start / End Page: 1 - 16 Identifier: ISSN: 1387-3806
CoNE: https://pure.mpg.de/cone/journals/resource/954926232412