English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
  A timing detector with pulsed high-voltage power supply for mass measurements at CSRe

Zhang, W., Tu X.L., X., Wang, M., Zhang, Y., Xu, H., Litvinov, Y. A., et al. (2014). A timing detector with pulsed high-voltage power supply for mass measurements at CSRe. Nuclear Instruments and Methods in Physics Research Section A, 755, 38-43. doi:10.1016/j.nima.2014.04.031.

Item is

Files

show Files

Locators

show
hide
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Zhang, W.1, 2, Author
Tu X.L., X.L.1, 3, 4, Author
Wang, M.1, Author
Zhang, Y.H.1, Author
Xu, H.S.1, Author
Litvinov, Yuri A.1, 3, 4, Author           
Blaum, Klaus4, Author           
Chen, X.C.1, 2, 3, Author
Hu, Z.G.1, Author
Huang, W.J.1, 2, Author
Ma, X.W.1, Author
Mao, R.S.1, Author
Mei, B.1, 3, 5, Author
Shuai, P.1, 6, Author
Sun, B.H.7, Author
Yamaguchi, T.8, Author
Xia, J.W.1, Author
Xiao, G.Q.1, Author
Xu, X.1, 2, Author
Yan, X.L.1, 2, Author
Yang, J.C.1, AuthorYuan, Y.J.1, AuthorZhou, X.H.1, AuthorZhao, H.W.1, AuthorZhao, T.C1, Author more..
Affiliations:
1Key Laboratory of High Precision Nuclear Spectroscopy and Center for Nuclear Matter Science, Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000,China, ou_persistent22              
2University of Chinese Academy of Sciences, Beijing 100049, China, ou_persistent22              
3GSI Helmholtzzentrum für Schwerionenforschung, 64291 Darmstadt, Germany, ou_persistent22              
4Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              
5J.W. Goethe University, 60323 Frankfurt, Germany, ou_persistent22              
6Research Center for Hadron Physics, National Laboratory of Heavy Ion Accelerator Facility in Lanzhou and University of Science and Technology of China, Hefei 230026, China, ou_persistent22              
7School of Physics and Nuclear Energy Engineering, Beihang University, Beijing 100191, China, ou_persistent22              
8Department of Physics, Saitama University, Saitama 338-8570, Japan, ou_persistent22              

Content

show
hide
Free keywords: -
 MPINP: Speichertechnik - Abteilung Blaum
 Abstract: Accuracy of nuclear mass measurements in storage rings depends critically on the accuracy with which the revolution times of stored ions can be obtained. In such experiments, micro-channel plates (MCP) are used as timing detectors. Due to large phase space of injected secondary beams, a large number of ions cannot be stored in the ring and is lost within the first few revolutions. However, these ions interact with the detector and can saturate the MCP and thus deteriorate its performance. In order to eliminate such effects, a fast, pulsed high-voltage power supply (PHVPS) has been employed which keeps the detector switched-off during the first few revolutions. The new detector setup was taken into operation at the Experimental Cooler-Storage-Ring CSRe in Lanzhou and resulted in a significant improvement of the detector amplitude and efficiency characteristics.

Details

show
hide
Language(s): eng - English
 Dates: 2014-04-23
 Publication Status: Published online
 Pages: 6
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1016/j.nima.2014.04.031
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Nuclear Instruments and Methods in Physics Research Section A
  Other : Nucl. Instrum. Methods Phys. Res. A
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: North-Holland
Pages: - Volume / Issue: 755 Sequence Number: - Start / End Page: 38 - 43 Identifier: ISSN: 0168-9002
CoNE: https://pure.mpg.de/cone/journals/resource/954925485716