日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細

  Dual-channel photoacoustic hygrometer for airborne measurements: background, calibration, laboratory and in-flight intercomparison tests

Tátrai, D., Bozóki, Z., Smit, H., Rolf, C., Spelten, N., Krämer, M., Filges, A., Gerbig, C., Gulyás, G., & Szabó, G. (2015). Dual-channel photoacoustic hygrometer for airborne measurements: background, calibration, laboratory and in-flight intercomparison tests. Atmospheric Measurement Techniques, 8, 33-42. doi:10.5194/amtd-7-6359-2014.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル
非表示: ファイル
:
BGC2060D.pdf (出版社版), 667KB
ファイルのパーマリンク:
https://hdl.handle.net/11858/00-001M-0000-0019-D252-A
ファイル名:
BGC2060D.pdf
説明:
Discussion paper
OA-Status:
閲覧制限:
公開
MIMEタイプ / チェックサム:
application/pdf / [MD5]
技術的なメタデータ:
著作権日付:
-
著作権情報:
-
:
BGC2060.pdf (出版社版), 919KB
ファイルのパーマリンク:
https://hdl.handle.net/11858/00-001M-0000-0024-65DA-8
ファイル名:
BGC2060.pdf
説明:
-
OA-Status:
閲覧制限:
公開
MIMEタイプ / チェックサム:
application/pdf / [MD5]
技術的なメタデータ:
著作権日付:
-
著作権情報:
-

関連URL

表示:
非表示:
説明:
OA
OA-Status:

作成者

表示:
非表示:
 作成者:
Tátrai, D., 著者
Bozóki, Z., 著者
Smit, H., 著者
Rolf, C., 著者
Spelten, N., 著者
Krämer, M., 著者
Filges, Annette1, 著者           
Gerbig, Christoph1, 著者           
Gulyás, G., 著者
Szabó, G., 著者
所属:
1Airborne Trace Gas Measurements and Mesoscale Modelling, Dr. habil. C. Gerbig, Department Biogeochemical Systems, Prof. M. Heimann, Max Planck Institute for Biogeochemistry, Max Planck Society, ou_1497784              

内容説明

表示:
非表示:
キーワード: -
 要旨: This paper describes a tunable diode laser based dual channel photoacoustic (PA) humidity measuring system called WaSul-Hygro primarily designed for aircraft based environment research. It is calibrated for total pressures and water vapor (WV) vol- 5 ume mixing ratios (VMRs) possible during airborne applications. WV VMR is calculated by using pressure dependent calibration curves and a cubic spline interpolation method. Coverage of the entire atmospheric humidity concentration range which might be encountered during airborne measurements is facilitated by applying an automated sensitivity mode switching algorithm. The calibrated PA system was validated through 10 laboratory and airborne inter-comparisons, which proved that the repeatability, the estimated accuracy and the response time of the system is 0.5 ppmV or 0.5% of the actual reading (whichever value is the greater), 5% of the actual reading within the VMR range of 1–12 000 ppmV and 2 s, respectively. The upper detection limit of the system is about 85 000 ppmV, limited only by condensation of water vapor on the walls 15 of the 318 K heated PA cells and inlet lines. The unique advantage of the presented system is its applicability for simultaneous water vapor and total water volume mixing ratio measurements.

資料詳細

表示:
非表示:
言語:
 日付: 2014-05-192014-06-262015
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): その他: BGC2060
DOI: 10.5194/amtd-7-6359-2014
ISSN: 1867-1381
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Atmospheric Measurement Techniques
  省略形 : AMT
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: Göttingen : European Geosciences Union, Copernicus
ページ: - 巻号: 8 通巻号: - 開始・終了ページ: 33 - 42 識別子(ISBN, ISSN, DOIなど): ISSN: 1867-1381
CoNE: https://pure.mpg.de/cone/journals/resource/1867-1381