Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

DATENSATZ AKTIONENEXPORT
  Fast T2 mapping with improved accuracy using undersampled spin-echo MRI and model-based reconstructions with a generating function.

Sumpf, T., Petrovic, A., Uecker, M., Knoll, F., & Frahm, J. (2014). Fast T2 mapping with improved accuracy using undersampled spin-echo MRI and model-based reconstructions with a generating function. IEEE Transactions on Medical Imaging, 33(12), 2213-2222. doi:10.1109/TMI.2014.2333370.

Item is

Dateien

einblenden: Dateien
ausblenden: Dateien
:
2078626.pdf (Verlagsversion), 2MB
 
Datei-Permalink:
-
Name:
2078626.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Eingeschränkt (UNKNOWN id 303; )
MIME-Typ / Prüfsumme:
application/pdf
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:
ausblenden:
Beschreibung:
-
OA-Status:

Urheber

einblenden:
ausblenden:
 Urheber:
Sumpf, T.1, Autor           
Petrovic, A., Autor
Uecker, M.1, Autor           
Knoll, F., Autor
Frahm, J.1, Autor           
Affiliations:
1Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society, ou_578634              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 Zusammenfassung: A model-based reconstruction technique for accelerated T2 mapping with improved accuracy is proposed using undersampled Cartesian spin-echo magnetic resonance imaging (MRI) data. The technique employs an advanced signal model for T2 relaxation that accounts for contributions from indirect echoes in a train of multiple spin echoes. An iterative solution of the nonlinear inverse reconstruction problem directly estimates spin-density and T2 maps from undersampled raw data. The algorithm is validated for simulated data as well as phantom and human brain MRI at 3T. The performance of the advanced model is compared to conventional pixel-based fitting of echo-time images from fully sampled data. The proposed method yields more accurate T2 values than the mono-exponential model and allows for retrospective undersampling factors of at least 6. Although limitations are observed for very long T2 relaxation times, respective reconstruction problems may be overcome by a gradient dampening approach. The analytical gradient of the utilized cost function is included as Appendix . The source code is made available to the community.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2014-06-232014-12
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1109/TMI.2014.2333370
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: IEEE Transactions on Medical Imaging
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: -
Seiten: - Band / Heft: 33 (12) Artikelnummer: - Start- / Endseite: 2213 - 2222 Identifikator: -