English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
  An evaluation of prospective motion correction (PMC) for high resolution quantitative MRI

Callaghan, M. F., Josephs, O., Herbst, M., Zaitsev, M., Todd, N., & Weiskopf, N. (2015). An evaluation of prospective motion correction (PMC) for high resolution quantitative MRI. Frontiers in Neuroscience, 9: 97. doi:10.3389/fnins.2015.00097.

Item is

Files

show Files
hide Files
:
Callaghan_Josephs_2015.pdf (Publisher version), 3MB
Name:
Callaghan_Josephs_2015.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show

Creators

show
hide
 Creators:
Callaghan, Martina F.1, Author
Josephs, Oliver1, Author
Herbst, Michael2, 3, Author
Zaitsev, Maxim2, Author
Todd, Nick1, Author
Weiskopf, Nikolaus1, Author           
Affiliations:
1Wellcome Trust Centre for Neuroimaging, University College London, United Kingdom, ou_persistent22              
2Department of Radiology, University Medical Center, Freiburg, Germany, ou_persistent22              
3Department of Medicine, John A. Burns School of Medicine, University of Hawaii, Honululu, HI, USA, ou_persistent22              

Content

show
hide
Free keywords: Prospective motion correction (PMC); Relaxometry; Quantitative; Multi-parameter mapping; MPM
 Abstract: Quantitative imaging aims to provide in vivo neuroimaging biomarkers with high research and diagnostic value that are sensitive to underlying tissue microstructure. In order to use these data to examine intra-cortical differences or to define boundaries between different myelo-architectural areas, high resolution data are required. The quality of such measurements is degraded in the presence of motion hindering insight into brain microstructure. Correction schemes are therefore vital for high resolution, whole brain coverage approaches that have long acquisition times and greater sensitivity to motion. Here we evaluate the use of prospective motion correction (PMC) via an optical tracking system to counter intra-scan motion in a high resolution (800 μm isotropic) multi-parameter mapping (MPM) protocol. Data were acquired on six volunteers using a 2 × 2 factorial design permuting the following conditions: PMC on/off and motion/no motion. In the presence of head motion, PMC-based motion correction considerably improved the quality of the maps as reflected by fewer visible artifacts and improved consistency. The precision of the maps, parameterized through the coefficient of variation in cortical sub-regions, showed improvements of 11–25% in the presence of deliberate head motion. Importantly, in the absence of motion the PMC system did not introduce extraneous artifacts into the quantitative maps. The PMC system based on optical tracking offers a robust approach to minimizing motion artifacts in quantitative anatomical imaging without extending scan times. Such a robust motion correction scheme is crucial in order to achieve the ultra-high resolution required of quantitative imaging for cutting edge in vivo histology applications.

Details

show
hide
Language(s): eng - English
 Dates: 2015-01-152015-03-062015-03-25
 Publication Status: Published online
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.3389/fnins.2015.00097
PMID: 25859178
PMC: PMC4373264
Other: eCollection 2015
 Degree: -

Event

show

Legal Case

show

Project information

show hide
Project name : Non-invasive in vivo histology in health and disease using Magnetic Resonance Imaging (MRI) / HMRI
Grant ID : 616905
Funding program : FP7 (ERC-2013-CoG)
Funding organization : European Research Council
Project name : -
Grant ID : -
Funding program : -
Funding organization : SLMS Captial Equipment Award (UCL)
Project name : -
Grant ID : -
Funding program : -
Funding organization : Alexander von Humboldt Foundation
Project name : -
Grant ID : 091593/Z/10/Z
Funding program : -
Funding organization : Wellcome Trust

Source 1

show
hide
Title: Frontiers in Neuroscience
  Other : Front Neurosci
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Lausanne, Switzerland : Frontiers Research Foundation
Pages: - Volume / Issue: 9 Sequence Number: 97 Start / End Page: - Identifier: ISSN: 1662-4548
ISSN: 1662-453X
CoNE: https://pure.mpg.de/cone/journals/resource/1662-4548