  New high-DQE imaging plate scanner using the reflected readout laser signal for noise corrections.

Bele, P., Ochs, R., Angert, I., & Schröder, R. R. (2000). New high-DQE imaging plate scanner using the reflected readout laser signal for noise corrections. Microscopy Research and Technique, 49(3), 281-291. doi:10.1002/(SICI)1097-0029(20000501)49:3<281:AID-JEMT6>3.0.CO;2-A.

Item is

Basic

show hide
Genre: Journal Article
Alternative Title : New high-DQE imaging plate scanner using the reflected readout laser signal for noise corrections.

Files

show Files
hide Files
:
MicroscopResTech_49_2000_281.pdf (Any fulltext), 369KB
 
File Permalink:
-
Name:
MicroscopResTech_49_2000_281.pdf
Description:
-
OA-Status:
Visibility:
Restricted (Max Planck Institute for Medical Research, MHMF; )
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Creators

show
hide
 Creators:
Bele, Petra1, Author           
Ochs, Rainer, Author
Angert, Isabel, Author           
Schröder, Rasmus R.1, Author           
Affiliations:
1Emeritus Group Biophysics, Max Planck Institute for Medical Research, Max Planck Society, ou_1497712              

Content

show
hide
Free keywords: Imaging Plate; scanner; reflected light; readout laser; high DQE; gain normalization; SNR correction
 Abstract: Imaging Plates (IPs) are in principle ideal electron detectors combining a large active layer area with a high sensitivity, linear dynamic range detection over 5 orders of magnitude. A moderate resolution and a decreasing detection quantum efficiency (DQE) for higher electron doses limit their use so far. The decrease of the DQE results from linear noise contributed by readout laser instabilities and inhomogeneities of the IP active layer. Here we present data on a new IP drum scanner prototype. This scanner combines twin channel amplification electronics with a new type of readout laser which allows a smaller readout focus and increased stability. The current nominal pixel size is 25 m, and the measured modulation transfer function (MTF) indicates that further reduction of the scanning step size down to pixel sizes in the range of 12−15 m should be possible. A unique feature of the new scanner is the simultaneous recording of the reflected readout laser light. The reflected light signal can be used for a posteriori alignment of repeated scans of one individual IP and for a correction of one part of the high spatial frequency noise contribution (reflected light correction). The posteriori alignment now allows an easy conventional gain normalization of the luminescence signal without using special markers on the IP. Both corrections lead to an increase of the DQE for high electron doses

Details

show
hide
Language(s): eng - English
 Dates: 1998-10-101999-02-162000-04-272000-05-01
 Publication Status: Issued
 Pages: 11
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Microscopy Research and Technique
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: New York, NY : Wiley-Liss
Pages: - Volume / Issue: 49 (3) Sequence Number: - Start / End Page: 281 - 291 Identifier: ISSN: 1059-910X
CoNE: https://pure.mpg.de/cone/journals/resource/954927686720