Importance: Despite significant advances in surgery, most surgical tools remain basic. Lasers provide a means of precise surgical ablation, but their clinical use has remained limited because of undesired thermal, ionizing, or acoustic stress effects leading to tissue injury. A novel ultrafast, nonionizing, picosecond infrared laser (PIRL) system has recently been developed and is capable, in theory, of ablation with negligible thermal or acoustic stress effects.
Objective: To measure and compare heat generation by means of thermography during ablation of ex vivo porcine skin by conventional microsecond-pulsed erbium:YAG (Er:YAG) laser and picosecond infrared laser (PIRL).
Design and Setting: This study was conducted in an optics laboratory and used a pretest-posttest experimental design comparing 2 methods of laser ablation of tissue with each sample acting as its own control.
Intervention: Ex vivo porcine skin was ablated in a 5-mm line pattern with both Er:YAG laser and PIRL at fluence levels marginally above ablation threshold (2 J/cm2 and 0.6 J/cm2, respectively).
Main Outcomes and Measures: Peaks and maxima of skin temperature rises were determined using a thermography camera. Means of peak temperature rises were compared using the paired sample t test. Ablation craters were assessed by means of digital microscopy.
Results: Mean peak rise in skin surface temperature for the Er:YAG laser and PIRL was 15.0°C and 1.68°C, respectively (P < .001). Maximum peak rise in skin surface temperature was 18.85°C for the Er:YAG laser and 2.05°C for the PIRL. Ablation craters were confirmed on digital microscopy.
Conclusions and Relevance: Picosecond infrared laser ablation results in negligible heat generation, considerably less than Er:YAG laser ablation, which confirms the potential of this novel technology in minimizing undesirable thermal injury associated with lasers currently in clinical use.