  16-channel bow tie antenna transceiver array for cardiac MR at 7.0 tesla.

Oezerdem, C., Winter, L., Graessl, A., Paul, K., Els, A., Weinberger, O., et al. (2016). 16-channel bow tie antenna transceiver array for cardiac MR at 7.0 tesla. Magnetic Resonance in Medicine, 75(6), 2553-2565. doi:10.1002/mrm.25840.

Item is

Files

show Files
hide Files
:
2179877.pdf (Publisher version), 9MB
 
File Permalink:
-
Name:
2179877.pdf
Description:
-
OA-Status:
Visibility:
Restricted (UNKNOWN id 303; )
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show
hide
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Oezerdem, C., Author
Winter, L., Author
Graessl, A., Author
Paul, K., Author
Els, A., Author
Weinberger, O., Author
Rieger, J., Author
Kuehne, A., Author
Dieringer, M., Author
Hezel, F., Author
Voit, D.1, Author           
Frahm, J.1, Author           
Niendorf, Thoralf, Author
Affiliations:
1Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society, ou_578634              

Content

show
hide
Free keywords: ultrahigh field MRI; cardiovascular MRI; electrical dipole; transceiver array; RF coil; parallel imaging
 Abstract: Purpose To design, evaluate, and apply a bow tie antenna transceiver radiofrequency (RF) coil array tailored for cardiac MRI at 7.0 Tesla (T). Methods The radiofrequency (RF) coil array comprises 16 building blocks each containing a bow tie shaped λ/2-dipole antenna. Numerical simulations were used for transmission field homogenization and RF safety validation. RF characteristics were examined in a phantom study. The array's suitability for high spatial resolution two-dimensional (2D) CINE imaging and for real time imaging of the heart was examined in a volunteer study. Results The arrays transmission fields and RF characteristics are suitable for cardiac MRI at 7.0T. The coil performance afforded a spatial resolution as good as (0.8 × 0.8 × 2.5) mm3 for segmented 2D CINE MRI at 7.0T which is by a factor of 12 superior versus standardized protocols used in clinical practice at 1.5T. The proposed transceiver array supports 1D acceleration factors of up to R = 6 without impairing image quality significantly. Conclusion The 16-channel bow tie antenna transceiver array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0 Tesla.

Details

show
hide
Language(s): eng - English
 Dates: 2015-07-172016-06
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1002/mrm.25840
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Magnetic Resonance in Medicine
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 75 (6) Sequence Number: - Start / End Page: 2553 - 2565 Identifier: -