English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
 
 
DownloadE-Mail
  Generation 3 programmable array microscope (PAM) for high speed large format optical sectioning in fluorescence.

de Vries, A. H. B., Cook, N., Kramer, S. C., Arndt-Jovin, D. J., & Jovin, T. M. (2015). Generation 3 programmable array microscope (PAM) for high speed large format optical sectioning in fluorescence. In M. R. Douglass, P. S. King, & B. L. Lee (Eds.), Emerging Digital Micromirror Device Based Systems and Applications VII (pp. 93760C-1-93760C-15). Bellingham, Wash.: SPIE. doi:10.1117/12.2076390.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
de Vries, A. H. B.1, Author           
Cook, N.1, Author           
Kramer, S. C.1, Author           
Arndt-Jovin, D. J.1, Author           
Jovin, T. M.1, Author           
Affiliations:
1Emeritus Group Laboratory of Cellular Dynamics, MPI for Biophysical Chemistry, Max Planck Society, ou_578629              

Content

show
hide
Free keywords: -
 Abstract: We report on the current version of the optical sectioning programmable array microscope (PAM) implemented with a single digital micro-mirror device (DMD) spatial light modulator utilized as a mask in both the fluorescence excitation and emission paths. The PAM incorporates structured illumination and structured detection operating in synchrony. A sequence of binary patterns of excitation light in high definition format (1920×1080 elements) is projected into the focal plane of the microscope at the 18 kHz binary frame rate of the Texas Instruments 1080p DMD. The resulting fluorescent emission is captured as two distinct signals: conjugate (c, ca. “on-focus”) consisting of light impinging on and deviated from the “on” elements of the DMD, and the non-conjugate (nc, ca. “out-of-focus”) light falling on and deviated from the “off” elements. The two distinct, deflected beams are optically filtered and detected either by two individual cameras or captured as adjacent images on a single camera after traversing an image combiner. The sectioned image is gained from a subtraction of the nc image from the c image, weighted in accordance with the pattern(s) used for illumination and detection and the relative exposure times of the cameras. The widefield image is given by the sum of the c and nc images. This procedure allows a high duty cycle (typically 25-50%) of on-elements in the excitation patterns and thus functions with low light intensities, preventing saturation and minimizing photobleaching of sensitive fluorophores. The corresponding acquisition speed is also very high, limited only by the bandwidth of the camera(s) (100 fps full frame with the sCMOS camera in current use) and the optical power of the light source (lasers, large area LEDs). In contrast to the static patterns typical of SIM systems, the programmable array allows optimization of the patterns (duty cycle, feature size and distribution), thus enabling a wide range of applications, ranging from patterned photobleaching, (e.g., FRAP, FLIP) and photoactivation, spatial superresolution, automated adaptive tracking and minimization of light exposure (MLE), and photolithography.

Details

show
hide
Language(s): eng - English
 Dates: 2015-03-102015
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1117/12.2076390
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Emerging Digital Micromirror Device Based Systems and Applications VII
Source Genre: Book
 Creator(s):
Douglass, M. R., Editor
King, P. S., Editor
Lee, B. L., Editor
Affiliations:
-
Publ. Info: Bellingham, Wash. : SPIE
Pages: 180 Volume / Issue: - Sequence Number: - Start / End Page: 93760C-1 - 93760C-15 Identifier: ISBN: 978-1-62841-466-0

Source 2

show
hide
Title: Proceedings of SPIE
Source Genre: Series
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 9376 Sequence Number: - Start / End Page: - Identifier: -