  Correction of inter-scan motion artifacts in quantitative R1 mapping by accounting for receive coil sensitivity effects

Papp, D., Callaghan, M. F., Meyer, H., Buckley, C., & Weiskopf, N. (2016). Correction of inter-scan motion artifacts in quantitative R1 mapping by accounting for receive coil sensitivity effects. Magnetic Resonance in Medicine, 76(5), 1478-1485. doi:10.1002/mrm.26058.

Item is

Basisdaten

einblenden: ausblenden:
Genre: Zeitschriftenartikel

Dateien

einblenden: Dateien
ausblenden: Dateien
:
Papp_2016.pdf (Verlagsversion), 522KB
Name:
Papp_2016.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Öffentlich
MIME-Typ / Prüfsumme:
application/pdf / [MD5]
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Papp, Daniel1, Autor
Callaghan, Martina F.1, Autor
Meyer, Heiko2, Autor
Buckley, Craig3, Autor
Weiskopf, Nikolaus1, 4, Autor           
Affiliations:
1Wellcome Trust Centre for Neuroimaging, University College London, United Kingdom, ou_persistent22              
2Siemens Sector Healthcare, Erlangen, Germany, ou_persistent22              
3Siemens Healthcare, Camberley, United Kingdom, ou_persistent22              
4Department Neurophysics (Weiskopf), MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_2205649              

Inhalt

einblenden:
ausblenden:
Schlagwörter: MPM; R1 mapping; VFA; Motion correction; Quantitative MRI; Receive sensitivity
 Zusammenfassung: PURPOSE:

Inter-scan motion causes differential receive field modulation between scans, leading to errors when they are combined to quantify MRI parameters. We present a robust and efficient method that accounts for inter-scan motion by removing this modulation before parameter quantification.
THEORY AND METHODS:

Five participants moved between two high-resolution structural scans acquired with different flip angles. Before each high-resolution scan, the effective relative sensitivity of the receive head coil was estimated by combining two rapid low-resolution scans acquired receiving on each of the body and head coils. All data were co-registered and sensitivity variations were removed from the high-resolution scans by division with the effective relative sensitivity. R1 maps with and without this correction were calculated and compared against reference maps unaffected by inter-scan motion.
RESULTS:

Even after coregistration, inter-scan motion significantly biased the R1 maps, leading to spurious variation in R1 in brain tissue and deviations with respect to a no-motion reference. The proposed correction scheme reduced the error to within the typical scan-rescan error observed in datasets unaffected by motion.
CONCLUSION:

Inter-scan motion negatively impacts the accuracy and precision of R1 mapping. We present a validated correction method that accounts for position-specific receive field modulation. Magn Reson Med, 2015. © 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2015-10-062015-06-252015-10-262015-11-262016-11-01
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1002/mrm.26058
PMID: 26608936
PMC: PMC5082493
Anderer: Epub 2015
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Magnetic Resonance in Medicine
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: New York : Wiley-Liss
Seiten: - Band / Heft: 76 (5) Artikelnummer: - Start- / Endseite: 1478 - 1485 Identifikator: ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149