日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細

  Site-Specific Cryo-focused Ion Beam Sample Preparation Guided by 3D Correlative Microscopy

Arnold, J., Mahamid, J., Lucic, V., de Marco, A., Fernandez, J.-J., Laugks, T., Mayer, T., Hyman, A. A., Baumeister, W., & Plitzko, J. M. (2016). Site-Specific Cryo-focused Ion Beam Sample Preparation Guided by 3D Correlative Microscopy. BIOPHYSICAL JOURNAL, 110(4), 860-869. doi:10.1016/j.bpj.2015.10.053.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Arnold, Jan1, 著者           
Mahamid, Julia1, 著者           
Lucic, Vladan1, 著者           
de Marco, Alex2, 著者
Fernandez, Jose-Jesus2, 著者
Laugks, Tim1, 著者           
Mayer, Tobias2, 著者
Hyman, Anthony A.2, 著者
Baumeister, Wolfgang1, 著者           
Plitzko, Jürgen M.1, 著者           
所属:
1Baumeister, Wolfgang / Molecular Structural Biology, Max Planck Institute of Biochemistry, Max Planck Society, ou_1565142              
2external, ou_persistent22              

内容説明

表示:
非表示:
キーワード: CRYOELECTRON TOMOGRAPHY; ELECTRON TOMOGRAPHY; LIGHT-MICROSCOPY; FLUORESCENCE MICROSCOPY; CELLS; SPECIMENS; PRECISION; MARKERS
 要旨: The development of cryo-focused ion beam (cryo-FIB) for the thinning of frozen-hydrated biological specimens enabled cryo-electron tomography (cryo-ET) analysis in unperturbed cells and tissues. However, the volume represented within a typical FIB lamella constitutes a small fraction of the biological specimen. Retaining low-abundance and dynamic subcellular structures or macromolecular assemblies within such limited volumes requires precise targeting of the FIB milling process. In this study, we present the development of a cryo-stage allowing for spinning-disk confocal light microscopy at cryogenic temperatures and describe the incorporation of the new hardware into existing workflows for cellular sample preparation by cryo-FIB. Introduction of fiducial markers and subsequent computation of three-dimensional coordinate transformations provide correlation between light microscopy and scanning electron microscopy/FIB. The correlative approach is employed to guide the FIB milling process of vitrified cellular samples and to capture specific structures, namely fluorescently labeled lipid droplets, in lamellas that are 300 nm thick. The correlation procedure is then applied to localize the fluorescently labeled structures in the transmission electron microscopy image of the lamella. This approach can be employed to navigate the acquisition of cryo-ET data within FIB-lamellas at specific locations, unambiguously identified by fluorescence microscopy.

資料詳細

表示:
非表示:
言語: eng - English
 日付: 2016
 出版の状態: 出版
 ページ: 10
 出版情報: -
 目次: -
 査読: 査読あり
 識別子(DOI, ISBNなど): ISI: 000370763800014
DOI: 10.1016/j.bpj.2015.10.053
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: BIOPHYSICAL JOURNAL
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: 600 TECHNOLOGY SQUARE, 5TH FLOOR, CAMBRIDGE, MA 02139 USA : CELL PRESS
ページ: - 巻号: 110 (4) 通巻号: - 開始・終了ページ: 860 - 869 識別子(ISBN, ISSN, DOIなど): ISSN: 0006-3495