Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

DATENSATZ AKTIONENEXPORT
  The PERCIVAL soft X-ray imager.

Wunderer, C. B., Marras, A., Bayer, M., Glaser, L., Göttlicher, P., Lange, S., et al. (2014). The PERCIVAL soft X-ray imager. Journal of Instrumentation, 9(3): C03056. doi:10.1088/1748-0221/9/03/C03056.

Item is

Dateien

einblenden: Dateien
ausblenden: Dateien
:
2359284.pdf (Verlagsversion), 4MB
 
Datei-Permalink:
-
Name:
2359284.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Eingeschränkt (UNKNOWN id 303; )
MIME-Typ / Prüfsumme:
application/pdf
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Wunderer, C. B., Autor
Marras, A., Autor
Bayer, M., Autor
Glaser, L.1, Autor           
Göttlicher, P., Autor
Lange, S., Autor
Pithan, F., Autor
Scholz, F., Autor
Seltmann, J., Autor
Shevyakov, I., Autor
Smoljanin, S., Autor
Viefhaus, J., Autor
Viti, M., Autor
Xia, Q., Autor
Zimmer, M., Autor
Klumpp, S., Autor
Gasiorek, P., Autor
Guerrini, N., Autor
Marsh, B., Autor
Sedgwick, I., Autor
Turchetta, R., AutorCautero, G., AutorFarina, S., AutorGiuressi, D., AutorMenk, R., AutorStebel, L., AutorYousef, H., AutorMarchal, J., AutorNicholls, T., AutorTartoni, N., AutorGraafsma, H., Autor mehr..
Affiliations:
1Research Group of Structural Dynamics of (Bio)Chemical Systems, MPI for Biophysical Chemistry, Max Planck Society, ou_578564              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 Zusammenfassung: With the increased brilliance of state-of-the-art Synchrotron radiation sources and the advent of Free Electron Lasers enabling revolutionary science with EUV to X-ray photons comes an urgent need for suitable photon imaging detectors. Requirements include high frame rates, very large dynamic range, single-photon counting capability with low probability of false positives, and (multi)-megapixels. PERCIVAL (``Pixelated Energy Resolving CMOS Imager, Versatile and Large'') is currently being developed by a collaboration of DESY, RAL, Elettra and DLS to address this need for the soft X-ray regime. PERCIVAL is a monolithic active pixel sensor (MAPS), i.e. based on CMOS technology. It will be back-thinned to access its primary energy range of 250 eV to 1 keV with target efficiencies above 90%. According to its preliminary specifications, the roughly 10 × 10 cm2, 3520 × 3710 pixel monolithic sensor will operate at frame rates up to 120 Hz (commensurate with most FELs) and use multiple gains within its 27 μm pixels to measure (e.g. at 500 eV) 1 to ~ 105 simultaneously-arriving photons. Currently, small-scale front-illuminated prototype systems (160 × 210 pixels) are undergoing detailed testing with visible-light as well as X-ray photons.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2014-03-312014-03
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1088/1748-0221/9/03/C03056
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Journal of Instrumentation
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: -
Seiten: 11 Band / Heft: 9 (3) Artikelnummer: C03056 Start- / Endseite: - Identifikator: -