  Highly sensitive superconducting circuits at ∼700 kHz with tunable quality factors for image-current detection of single trapped antiprotons

Nagahama, H., Schneider Georg, G., Mooser, A., Smorra, C., Sellner, S., Harrington, J., et al. (2016). Highly sensitive superconducting circuits at ∼700 kHz with tunable quality factors for image-current detection of single trapped antiprotons. Review of Scientific Instruments, 87(11): 113305. doi:10.1063/1.4967493.

Item is

Files

show Files

Locators

show
hide
Locator:
http://dx.doi.org/10.1063/1.4967493 (Publisher version)
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Nagahama, H.1, 2, Author
Schneider Georg, G.1, 3, Author
Mooser, A.1, Author
Smorra, C.1, 4, Author
Sellner, S.1, Author
Harrington, J.5, Author           
Higuchi, T.1, 2, Author
Borchert, M.6, Author
Tanaka, T.1, 2, Author
Besirli, M.1, Author
Blaum, Klaus5, Author           
Matsuda, Y.2, Author
Ospelkaus, C.6, 7, Author
Quint, W.8, Author
Walz, J.3, 9, Author
Yamazaki, Y.10, Author
Ulmer, S.1, Author
Affiliations:
1RIKEN, Ulmer Initiative Research Unit, 2-1 Hirosawa, Wako, Saitama 351-0198, Japan, ou_persistent22              
2Graduate School of Arts and Sciences, University of Tokyo, Tokyo 153-8902, Japan, ou_persistent22              
3Institut für Physik, Johannes Gutenberg-Universität, 55099 Mainz, Germany, ou_persistent22              
4CERN, 1211 Geneva, Switzerland, ou_persistent22              
5Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              
6Institut für Quantenoptik, Leibniz Universität Hannover, Welfengarten 1, 30167 Hannover, Germany, ou_persistent22              
7Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany, ou_persistent22              
8GSI - Helmholtzzentrum für Schwerionenforschung GmbH, 64291 Darmstadt, Germany, ou_persistent22              
9Helmholtz-Institut Mainz, 55099 Mainz, Germany, ou_persistent22              
100RIKEN, Atomic Physics Research Unit, 2-1 Hirosawa, Wako, Saitama 351-0198, Japan, ou_persistent22              

Content

show
hide
Free keywords: -
 MPINP: Speichertechnik - Abteilung Blaum
 Abstract: We developed highly sensitive image-current detection systems based on superconducting toroidal coils and ultra-low noise amplifiers for non-destructive measurements of the axial frequencies (550–800 kHz) of single antiprotons stored in a cryogenic multi-Penning-trap system. The unloaded superconducting tuned circuits show quality factors of up to 500 000, which corresponds to a factor of 10 improvement compared to our previously used solenoidal designs. Connected to ultra-low noise amplifiers and the trap system, signal-to-noise-ratios of 30 dB at quality factors of >20 000 are achieved. In addition, we have developed a superconducting switch which allows continuous tuning of the detector’s quality factor and to sensitively tune the particle-detector interaction. This allowed us to improve frequency resolution at constant averaging time, which is crucial for single antiproton spin-transition spectroscopy experiments, as well as improved measurements of the proton-to-antiproton charge-to-mass ratio.

Details

show
hide
Language(s):
 Dates: 2016-11-15
 Publication Status: Published online
 Pages: 6
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1063/1.4967493
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Review of Scientific Instruments
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Melville, NY : AIP Publishing
Pages: - Volume / Issue: 87 (11) Sequence Number: 113305 Start / End Page: - Identifier: ISSN: 0034-6748
CoNE: https://pure.mpg.de/cone/journals/resource/991042742033452