hide
Free keywords:
-
Abstract:
222Rn is a major source of background in many rare-event experiments such as the
XENON1T dark matter search. The noble gas radon is created by radioactive decay inside
all detector materials and emanates into the sensitive liquid xenon target disabling
any detector shielding. Subsequent beta-decays of radon progenies are the dominant
source of background in the XENON1T experiment. In order to mitigate radon induced
background the detector's construction materials have been selected according to dedicated
222Rn emanation measurements. In the first part of this thesis, we summarize
the results of the XENON1T radon screening campaign and present the measurement
of the integral radon emanation rate of the fully assembled detector. The development
of a radon removal system which continuously purifies the liquid xenon target from the
emanated radon is the topic of the second part of this thesis. In order to demonstrate
the suitability of cryogenic distillation as a technique to separate radon from xenon, we
developed an experimental setup to measure the depletion of radon in xenon boil-off
gas after a single distillation step. In the last part of the thesis, we demonstrate the
operation of a radon removal system for the XENON100 experiment. For this first test
employing a running dark matter detector, we integrated a multiple stage, cryogenic
distillation column in the XENON100 gas purification loop. From the evolution of the
radon concentration in XENON100, we investigate the distillation column's radon removal
capability and discuss the design and application of a radon removal system for
XENON1T and the upcoming XENONnT experiment.