Purpose
The study compares glycosaminoglycan chemical exchange saturation transfer (gagCEST) imaging of intervertebral discs corrected for solely B0 inhomogeneities or both B0 and B1 inhomogeneities.
Methods
Lumbar intervertebral discs of 20 volunteers were examined with T2-weighted and gagCEST imaging. Field inhomogeneity correction was performed with B0 correction only and with correction of both B0 and B1. GagCEST effects measured by the asymmetric magnetization transfer ratio (MTRasym) and signal-to-noise ratio (SNR) were compared between both methods.
Results
Significant higher MTRasym and SNR values were obtained in the nucleus pulposus using B0 and B1 correction compared with B0-corrected gagCEST. The GagCEST effect was significantly different in the nucleus pulposus compared with the annulus fibrosus for both methods.
Conclusion
The B0 and B1 field inhomogeneity correction method leads to an improved quality of gagCEST imaging in IVDs compared with only B0 correction.