English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
 
 
DownloadE-Mail
  Closed-cycle, low-vibration 4 K cryostat for ion traps and other applications

Micke, P., Stark, J., King, S. A., Leopold, T., Pfeifer, T., Schmoeger, L., et al. (2019). Closed-cycle, low-vibration 4 K cryostat for ion traps and other applications. Review of Scientific Instruments, 90(6): 065104. doi:10.1063/1.5088593.

Item is

Files

show Files
hide Files
:
1901.03630.pdf (Preprint), 10MB
Name:
1901.03630.pdf
Description:
File downloaded from arXiv at 2019-06-11 13:32
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-

Locators

show
hide
Locator:
https://doi.org/10.1063/1.5088593 (Publisher version)
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Micke, P.1, Author           
Stark, J.1, Author           
King, S. A., Author
Leopold, T., Author
Pfeifer, T.1, Author           
Schmoeger, Lisa1, Author           
Schwarz, M.1, Author           
Spieß, Lukas1, Author           
Schmidt, P. O., Author
Crespo López Urrutia, J. R.1, Author           
Affiliations:
1Division Prof. Dr. Thomas Pfeifer, MPI for Nuclear Physics, Max Planck Society, ou_2025284              

Content

show
hide
Free keywords: Physics, Atomic Physics, physics.atom-ph,physics.app-ph, Physics, Instrumentation and Detectors, physics.ins-det,Quantum Physics, quant-ph
 Abstract: In vacuo cryogenic environments are ideal for applications requiring both low temperatures and extremely low particle densities. This enables reaching long storage and coherence times, for example, in ion traps, essential requirements for experiments with highly charged ions, quantum computation, and optical clocks. We have developed a novel cryostat continuously refrigerated with a pulse-tube cryocooler and providing the lowest vibration level reported for such a closed-cycle system with 1 W cooling power for a <5 K experiment. A decoupling system suppresses vibrations from the cryocooler by three orders of magnitude down to a level of 10 nm peak amplitudes in the horizontal plane. Heat loads of about 40 W (at 45 K) and 1 W (at 4 K) are transferred from an experimental chamber, mounted on an optical table, to the cryocooler through a vacuum-insulated massive 120 kg inertial copper pendulum. The 1.4 m long pendulum allows installation of the cryocooler in a separate, acoustically isolated machine room. At the experimental chamber, we measured the residual vibrations using an interferometric setup. The positioning of the 4 K elements is reproduced to better than a few micrometer after a full thermal cycle to room temperature. Extreme high vacuum on the 10−15 mbar level is achieved. In collaboration with the Max-Planck-Institut für Kernphysik, such a setup is now in operation at the Physikalisch-Technische Bundesanstalt for a next-generation optical clock experiment using highly charged ions.

Details

show
hide
Language(s):
 Dates: 2019-06-05
 Publication Status: Published online
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: arXiv: 1901.03630
DOI: 10.1063/1.5088593
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Review of Scientific Instruments
  Abbreviation : Rev. Sci. Instrum.
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Melville, NY : AIP Publishing
Pages: - Volume / Issue: 90 (6) Sequence Number: 065104 Start / End Page: - Identifier: ISSN: 0034-6748
CoNE: https://pure.mpg.de/cone/journals/resource/991042742033452