  Environmentally-induced systematic effects at the high-precision mass spectrometer PENTATRAP

Kromer, K. (2019). Environmentally-induced systematic effects at the high-precision mass spectrometer PENTATRAP. Master Thesis, Ruprecht-Karls-Universität, Heidelberg.

Item is

Basisdaten

einblenden: ausblenden:
Genre: Hochschulschrift

Dateien

einblenden: Dateien
ausblenden: Dateien
:
Masterthesis_KathrinKromer.pdf (beliebiger Volltext), 10MB
Name:
Masterthesis_KathrinKromer.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Öffentlich
MIME-Typ / Prüfsumme:
application/pdf / [MD5]
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Kromer, Kathrin1, Autor           
Affiliations:
1Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 MPINP: Präzisionsexperimente - Abteilung Blaum
 Zusammenfassung: PENTATRAP is a mass spectrometer consisting of five Penning traps aiming for mass-ratio measurements of long-lived, highly-charged ions to a relative precision of below 10−11 by determining the cyclotron frequencies of trapped ions in a strong magnetic field. At this level of precision, mass-ratio determinations contribute, among others, to neutrino physics and a direct test of special relativity. In order to reach this goal, the systematic uncertainties of the cyclotron frequency measurement originating from environmental influences, including magnetic field, pressure, and temperature fluctuations, have to be precisely known and avoided. In the framework of this thesis, the main sources of magnetic field changes have been identified and an active magnetic field stabilization system was constructed and tested. The temperature was optimized to a stability of 20 mK/h and the temperature dependence of the voltage source for the trapping potentials of the measurement traps was determined to be 2.43(14) · 10−6/K and less. A helium-level and cold-bore
pressure stabilization system has been installed and successfully tested, reducing the fluctuations of the magnetic trapping field to a nearly constant relative drift of −2 · 10−10/h. With the achieved control over the environmental influences the first cyclotron frequency measurements of 187Re/187Re to a relative precision of 8 · 10−12 have been carried out.

Details

einblenden:
ausblenden:
Sprache(n):
 Datum: 2019
 Publikationsstatus: Angenommen
 Seiten: V, 58 S. : Ill., graph. Darst.
 Ort, Verlag, Ausgabe: Heidelberg : Ruprecht-Karls-Universität
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: -
 Art des Abschluß: Master

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle

einblenden: