ausblenden:
Schlagwörter:
-
Zusammenfassung:
Robust and fast measurements of the transmit field strength is of great importance particularly in parallel transmit applications at ultra-high field. In this work, a pre-saturation TurboFLASH B1+ mapping sequence was optimized for 3-dimensional single-shot acquisition with spiral-centric reordering. Improved SNR and reduced artifacts were achieved by using a variable flip-angle readout. The ability of the proposed sampling scheme to perform fast whole-brain B1+ mapping with low SAR was demonstrated in phantom and in-vivo experiments at 9.4 T. The obtained B1+ maps were comparable to those obtained with the conventional 2D-SatTFL approach but required significantly less scan time.