  Quantitative backscattered electron imaging of bone using a thermionic or a field emission electron source

Hartmann, M. A., Blouin, S., Misof, B. M., Fratzl-Zelman, N., Roschger, P., Berzlanovich, A., et al. (2021). Quantitative backscattered electron imaging of bone using a thermionic or a field emission electron source. Calcified Tissue International, 109(2), 190-202. doi:10.1007/s00223-021-00832-5.

Item is

Basic

show hide
Genre: Journal Article

Files

show Files
hide Files
:
Article.pdf (Publisher version), 5MB
Name:
Article.pdf
Description:
-
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-

Locators

show

Creators

show
hide
 Creators:
Hartmann, Markus A., Author
Blouin, Stéphane, Author
Misof, Barbara M., Author
Fratzl-Zelman, Nadja, Author
Roschger, Paul, Author
Berzlanovich, Andrea, Author
Gruber, Gerlinde M., Author
Brugger, Peter C., Author
Zwerina, Jochen, Author
Fratzl, Peter1, Author              
Affiliations:
1Peter Fratzl, Biomaterialien, Max Planck Institute of Colloids and Interfaces, Max Planck Society, ou_1863294              

Content

show
hide
Free keywords: Quantitative backscattered electron imaging; Bone mineralization density distribution; Adult human bone
 Abstract: Quantitative backscattered electron imaging is an established method to map mineral content distributions in bone and to determine the bone mineralization density distribution (BMDD). The method we applied was initially validated for a scanning electron microscope (SEM) equipped with a tungsten hairpin cathode (thermionic electron emission) under strongly defined settings of SEM parameters. For several reasons, it would be interesting to migrate the technique to a SEM with a field emission electron source (FE-SEM), which, however, would require to work with different SEM parameter settings as have been validated for DSM 962. The FE-SEM has a much better spatial resolution based on an electron source size in the order of several 100 nanometers, corresponding to an about 105 to 106 times smaller source area compared to thermionic sources. In the present work, we compare BMDD between these two types of instruments in order to further validate the methodology. We show that a transition to higher pixel resolution (1.76, 0.88, and 0.57 μm) results in shifts of the BMDD peak and BMDD width to higher values. Further the inter-device reproducibility of the mean calcium content shows a difference of up to 1 wt% Ca, while the technical variance of each device can be reduced to ±0.17 wt% Ca. Bearing in mind that shifts in calcium levels due to diseases, e.g., high turnover osteoporosis, are often in the range of 1 wt% Ca, both the bone samples of the patients as well as the control samples have to be measured on the same SEM device. Therefore, we also constructed new reference BMDD curves for adults to be used for FE-SEM data comparison.

Details

show
hide
Language(s): eng - English
 Dates: 2021-04-102021
 Publication Status: Published in print
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.1007/s00223-021-00832-5
PMID: 0608
Other: M:\BM-Publications\2021\HartmannCTI_Quantitative
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Calcified Tissue International
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: New York : Springer-Verlag
Pages: - Volume / Issue: 109 (2) Sequence Number: - Start / End Page: 190 - 202 Identifier: ISSN: 0171-967X