hide
Free keywords:
-
Abstract:
A compact retarding-potential Mott polarimeter combined with microsphere plates (MSP) as
electron detectors was built to perform spin-resolved time-of-flight electron spectroscopy. The comparison of the performance of MSP and channeltron detectors shows that the MSP detector has a better time resolution but a lower efficiency. The overall time resolution of the system was determined to be 350 ps using synchrotron radiation pulses.