  Commissioning and performance of TITAN's Multiple-Reflection Time-of-Flight Mass-Spectrometer and isobar separator

Reiter, M. P., San Andres, S. A., Bergmann, J., Dickel, T., Dilling, J., Jacobs, A., et al. (2021). Commissioning and performance of TITAN's Multiple-Reflection Time-of-Flight Mass-Spectrometer and isobar separator. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1018: 165823. doi:10.1016/j.nima.2021.165823.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Reiter, M. P.1, Author
San Andres, S. Ayet1, Author
Bergmann, J.1, Author
Dickel, T.1, Author
Dilling, J.1, Author
Jacobs, A.1, Author
Kwiatkowski, A. A.1, Author
Plass, W. R.1, Author
Scheidenberger, C.1, Author
Short, D.1, Author
Will, C.1, Author
Babcock, C.1, Author
Dunling, E.1, Author
Finlay, A.1, Author
Hornung, C.1, Author
Jesch, C.1, Author
Klawitter, Renee2, Author           
Kootte, B.1, Author
Lascar, D.1, Author
Leistenschneider, E.1, Author
Murboeck, T.1, AuthorPaul, S. F.1, AuthorYavor, M.1, Author more..
Affiliations:
1external, ou_persistent22              
2Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              

Content

show
hide
Free keywords: -
 MPINP: Speichertechnik - Abteilung Blaum
 Abstract: TRIUMF's Ion Trap for Atomic and Nuclear science (TITAN) has specialized
in fast Penning Trap mass spectrometry of very short-lived radioactive
isotopes. The facility has been upgraded with a Multiple-Reflection
Time-of-Flight Mass Spectrometer (MR-TOF-MS) to continue its quest
towards more exotic nuclides, which are critical for our understanding
of nuclear structure effects far from the valley of beta stability and
for the nucleosynthesis of heavy elements in explosive astrophysical
environments. In this publication, we discuss the implementation,
operation and performance of TITAN's MR-TOF-MS as a stand-alone
high-precision mass spectrometer and as an isobar separator. By using
the novel mass-selective re-trapping technique for the isobar
separation, the MR-TOF-MS can consecutively perform separation and mass
measurement of the same ion population, acting as its own isobar
separator. The device boosts the dynamic range and reach of the TITAN
facility by several orders of magnitude. The MR-TOF-MS reaches a high
mass resolving power (m/Delta m similar to 400 000), high precision and
mass accuracy (delta m/m < 10(-7)), is fast (common cycle time 20 ms),
shows high sensitivity and very large dynamic range (ion of interest to
contaminant ratios of up to 1 to 10(6)).

Details

show
hide
Language(s):
 Dates: 2021-12-01
 Publication Status: Published online
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Amsterdam : Elsevier B.V.
Pages: - Volume / Issue: 1018 Sequence Number: 165823 Start / End Page: - Identifier: ISSN: 0168-9002
CoNE: https://pure.mpg.de/cone/journals/resource/954925485716