Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

DATENSATZ AKTIONENEXPORT
  Characterization of an ultra-stable voltage supply and implementation of sympathetic laser cooling for the 3He2+ g-factor measurement

Kaiser, A. V. (2022). Characterization of an ultra-stable voltage supply and implementation of sympathetic laser cooling for the 3He2+ g-factor measurement. Master Thesis, Ruprecht-Karls-Universität, Heidelberg.

Item is

Basisdaten

einblenden: ausblenden:
Genre: Hochschulschrift

Dateien

einblenden: Dateien
ausblenden: Dateien
:
2022_Kaiser_Master.pdf (beliebiger Volltext), 15MB
Name:
2022_Kaiser_Master.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Öffentlich
MIME-Typ / Prüfsumme:
application/pdf / [MD5]
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Kaiser, Annabelle Valerie1, Autor           
Affiliations:
1Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 MPINP: Präzisionsexperimente - Abteilung Blaum
 Zusammenfassung: The high-precision measurement of the nuclear g-factor of a single 3He2+stored in
a Penning trap enables most accurate magnetometry using He nuclear magnetic
resonance probes. The g-factor measurement requires the detection of an axial
frequency shift caused by a spin flip. Due to the large mass and small magnetic
moment of 3He2+compared to the proton and electron, this frequency shift is
on the order of background frequency fluctuations, complicating the spin-state
determination. The signal-to-noise ratio can be enhanced by implementing two
upgrades to the Penning-trap setup. First, the electrode that generates the trap
potential must be supplied with an ultra-stable voltage source that must be tunable
in the range of a few hundred mV. At low electrode voltages (≤ 500mV),
the commercially available voltage source UM1-14 allows spin-flip detection with
99.2+0,4−2,7% fidelity. However, for larger electrode voltages the UM1-14 is no longer
suitable. Instead, a voltage source based on Josephson junctions was applied in
the context of this thesis and gives a spin-flip detection fidelity of 98+1
>−3%. Second,
the axial fluctuations are reduced by cooling the ion to low energies. This
is achieved by sympathetic laser cooling of 3He2+coupled to laser cooled 9Be+
After successful laser alignment, 9Be+was loaded into the trap and laser cooled.
An upper limit of the achieved 9Be+axial temperature is presently at 4.2K,
which is far above the Doppler cooling limit of 0.5mK.

Details

einblenden:
ausblenden:
Sprache(n):
 Datum: 2022
 Publikationsstatus: Angenommen
 Seiten: 72 S.
 Ort, Verlag, Ausgabe: Heidelberg : Ruprecht-Karls-Universität
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: -
 Art des Abschluß: Master

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle

einblenden: