  Determination of Temporal Stability and Instrument Performance of an airborne QCLAS via Allan-Werle-plots

Röder, L. L., Ort, L. M., Lelieveld, J., & Fischer, H. (2023). Determination of Temporal Stability and Instrument Performance of an airborne QCLAS via Allan-Werle-plots. Research Square. doi:10.21203/rs.3.rs-3619758/v1.

Item is

Files

show Files

Locators

show
hide
Description:
-
OA-Status:
Gold

Creators

show
hide
 Creators:
Röder, Lenard L.1, Author           
Ort, Linda M.1, Author           
Lelieveld, Jos1, Author           
Fischer, Horst1, Author           
Affiliations:
1Atmospheric Chemistry, Max Planck Institute for Chemistry, Max Planck Society, ou_1826285              

Content

show
hide
Free keywords: -
 Abstract: Allan-Werle-plots are an established tool in infrared absorption spectroscopy to quantify temporal stability, maximum integration time and best achievable precision of a measurement instrument. In field measurements aboard a moving platform, however, long integration times reduce time resolution and smooth atmospheric variability. A high accuracy and time resolution are necessary as well as an appropriate estimate of the measurement uncertainty. In this study, Allan-Werle-plots of calibration gas measurements are studied to analyze the temporal characteristics of a Quantum Cascade Laser Absorption Spectrometer (QCLAS) instrument for airborne operation. Via least-squares fitting the individual noise contributions can be quantified and different dominant regimes can be identified. Through simulation of data according to the characteristics from the Allan-Werle-plot, the effects of selected intervals between in-flight calibrations can be analyzed. An interval of 30 min is found sufficient for successful drift correction during ground operation. The linear interpolation of the sensitivity increases the accuracy and lowers the measurement uncertainty from 1.1 % to 0.2 %. Airborne operation yields similar results during segments of stable flight but additional flicker and sinusoidal contributions. Simulations verify an appropriate interval of 30 min in airborne operation. The expected airborne measurement uncertainty is in the range 0.8 % to 2.4 %.

Details

show
hide
Language(s): eng - English
 Dates: 2023-11-19
 Publication Status: Published online
 Pages: 16
 Publishing info: -
 Table of Contents: Under review for Applied Physics B
 Rev. Type: No review
 Identifiers: DOI: 10.21203/rs.3.rs-3619758/v1
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Research Square
Source Genre: Web Page
 Creator(s):
Affiliations:
Publ. Info: -