  Intra-scan RF power amplifier drift correction

Aghaeifar, A., Bosch, D., Heule, R., Williams, S., Ehses, P., Mauconduit, F., et al. (2024). Intra-scan RF power amplifier drift correction. Magnetic Resonance in Medicine, Epub ahead. doi:10.1002/mrm.30078.

Item is

Files

show Files

Locators

show
hide
Description:
-
OA-Status:
Not specified

Creators

show
hide
 Creators:
Aghaeifar, A1, Author                 
Bosch, D1, Author                 
Heule, R1, Author                 
Williams, S, Author
Ehses, P, Author                 
Mauconduit, F, Author
Scheffler, K1, Author                 
Affiliations:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

Content

show
hide
Free keywords: -
 Abstract: Purpose: The drift in radiofrequency (RF) power amplifiers (RFPAs) is assessed and several contributing factors are investigated. Two approaches for prospective correction of drift are proposed and their effectiveness is evaluated.
Methods: RFPA drift assessment encompasses both intra-pulse and inter-pulse drift analyses. Scan protocols with varying flip angle (FA), RF length, and pulse repetition time (TR) are used to gauge the influence of these parameters on drift. Directional couplers (DICOs) monitor the forward waveforms of the RFPA outputs. DICOs data is stored for evaluation, allowing calculation of correction factors to adjust RFPAs' transmit voltage. Two correction methods, predictive and run-time, are employed: predictive correction necessitates a calibration scan, while run-time correction calculates factors during the ongoing scan.
Results: RFPA drift is indeed influenced by the RF duty-cycle, and in the cases examined with a maximum duty-cycle of 66%, the potential drift is approximately 41% or 15%, depending on the specific RFPA revision. Notably, in low transmit voltage scenarios, FA has minimal impact on RFPA drift. The application of predictive and run-time drift correction techniques effectively reduces the average drift from 10.0% to less than 1%, resulting in enhanced MR signal stability.
Conclusion: Utilizing DICO recordings and implementing a feedback mechanism enable the prospective correction of RFPA drift. Having a calibration scan, predictive correction can be utilized with fewer complexity; for enhanced performance, a run-time approach can be employed.

Details

show
hide
Language(s):
 Dates: 2024-03
 Publication Status: Published online
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.1002/mrm.30078
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Magnetic Resonance in Medicine
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: New York : Wiley-Liss
Pages: - Volume / Issue: Epub ahead Sequence Number: - Start / End Page: - Identifier: ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149