hide
Free keywords:
-
Abstract:
Motivation: Conventional quantitative CEST experiments require the additional determination of $$$T_1$$$. CEST sequences include $$$T_1$$$ relaxation periods that can be exploited to estimate $$$T_1$$$ simultaneously to the CEST parameters.
Goal(s): The development of a technique for efficient simultaneous quantification of CEST contrast and $$$T_1$$$.
Approach: Integration of a FLASH acquisition train after conventional CEST saturations. Simultaneous CEST contrast and $$$T_1$$$ fitting is achieved by extending the Lorentzian model of the CEST spectrum with a Look-Locker model for the FLASH readout.
Results: The proof-of-concept was implemented, and first results are demonstrated in a two pools phantom and a four pool in vivo study.
Impact: The presented technique enables calculating $$$T_1$$$ corrected quantitative CEST results from one measurement. This simplifies the application of the apparent exchange-dependent relaxation ($$$MTR_{AREX}$$$), the quasi-steady-state$$$\;$$$contrast and other CEST metrics, which require a $$$T_1$$$ map, making them more accessible.