  An assessment of eddy current sensitivity and correction in single-shot diffusion-weighted imaging

Koch, M. A., & Norris, D. G. (2000). An assessment of eddy current sensitivity and correction in single-shot diffusion-weighted imaging. Physics in Medicine and Biology, 45(12), 3821-3832.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Koch, Martin A.1, Author           
Norris, David G.1, Author           
Affiliations:
1MPI of Cognitive Neuroscience (Leipzig, -2003), The Prior Institutes, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634574              

Content

show
hide
Free keywords: -
 Abstract: Artefacts caused by eddy currents are a major problem in diffusion weighted imaging. This is particularly acute in experiments in which a number of images with differing degrees of diffusion weighting and/or differently oriented diffusion-weighting gradients need to be combined. The echo-planar imaging sequence is particularly sensitive to the effects of residual eddy currents, especially due to the low bandwidth in the phase-encoding direction. Two published schemes are investigated regarding the effectiveness of eddy current correction. That of Jezzard et al (1998 Magn. Reson. Med. 39 801-12) requires the acquisition of additional experimental data in order to perform a post-acquisition correction, whereas that of Wider et al (1994 J. Magn. Reson. A 108 255-8) attempts to reduce the eddy currents directly. It is found that the latter experiment gives a somewhat superior performance and a combination of the two approaches results in an almost complete elimination of artefact. An alternative single-shot imaging experiment to echo-planar imaging is given by sequences based on fast spin-echo methods, which should be insensitive to the effects of constant eddy currents. It is shown that the intrinsic eddy-current-related artefact level in such experiments is indeed low, residual artefacts being attributed to eddy current decay during the echo train. In situations of poor main magnetic field homogeneity or large eddy currents such sequences may be gainfully used instead of echo-planar imaging.

Details

show
hide
Language(s): eng - English
 Dates: 2000
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: eDoc: 239006
ISI: 000165934800023
Other: P6566
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Physics in Medicine and Biology
  Other : Phys. Med. Biol.
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: London? : IOP Pub.
Pages: - Volume / Issue: 45 (12) Sequence Number: - Start / End Page: 3821 - 3832 Identifier: ISSN: 0031-9155
CoNE: https://pure.mpg.de/cone/journals/resource/954925433410