  10K Ring Electrode Trap - Tandem Mass Spectrometer for Infrared Spectroscopy of Mass Selected Ions

Goebbert, D. J., Meijer, G., & Asmis, K. R. (2009). 10K Ring Electrode Trap - Tandem Mass Spectrometer for Infrared Spectroscopy of Mass Selected Ions. AIP Conference Proceedings, 1104, 22-29.

Item is

Files

show Files
hide Files
:
399006(Goebbert).pdf (Preprint), 996KB
Name:
399006(Goebbert).pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-
:
736244CTA.pdf (Copyright transfer agreement), 2MB
 
File Permalink:
-
Name:
736244CTA.pdf
Description:
-
OA-Status:
Visibility:
Private
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-
:
736244.pdf (Correspondence), 2MB
 
File Permalink:
-
Name:
736244.pdf
Description:
-
OA-Status:
Visibility:
Private
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show

Creators

show
hide
 Creators:
Goebbert, Daniel J.1, Author           
Meijer, Gerard1, Author           
Asmis, Knut R.1, Author           
Affiliations:
1Molecular Physics, Fritz Haber Institute, Max Planck Society, ou_634545              

Content

show
hide
Free keywords: Ring electrode ion trap, infrared photodissociation spectroscopy, mass-selected ions; MOLECULAR-IONS; CLUSTER IONS; RESOLUTION; SPECTRA
 Abstract: A novel instrumental setup for measuring infrared photodissociation spectra of buffer gas cooled, mass-selected ions is described and tested. It combines a cryogenically cooled, linear radio frequency ion trap with a tandem mass spectrometer, optimally coupling continuous ion sources to pulsed laser experiments. The use of six independently adjustable DC potentials superimposed over the trapping radio frequency field provides control over the ion distribution within, as well as the kinetic energy distribution of the ions extracted from the ion trap. The scheme allows focusing the ions in space and time, such that they can be optimally irradiated by a pulsed, widely tunable infrared photodissociation laser. Ion intensities are monitored with a time-of-flight mass spectrometer mounted orthogonally to the ion trap axis.

Details

show
hide
Language(s): eng - English
 Dates: 2009-03-17
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1063/1.3115605
 Degree: -

Event

show
hide
Title: 4th International Conference on Laser Probing LAP 2008
Place of Event: Nagoya, Japan
Start-/End Date: 2008-10-06 - 2008-10-10

Legal Case

show

Project information

show

Source 1

show
hide
Title: AIP Conference Proceedings
  Alternative Title : AIP Conf. Proc.
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: 2 HUNMELVILLE, NY 11747-4501 USA : American Institute of Physics
Pages: - Volume / Issue: 1104 Sequence Number: - Start / End Page: 22 - 29 Identifier: ISSN: 0094-243X
ISBN: 978-0-7354-0640-7