Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

DATENSATZ AKTIONENEXPORT
  Effects of air susceptibility on proton resonance frequency MR thermometry

Streicher, M., Schäfer, A., Reimer, E., Dhital, B., Trampel, R., Ivanov, D., et al. (2012). Effects of air susceptibility on proton resonance frequency MR thermometry. Magnetic Resonance Materials in Physics, Biology and Medicine, 25(1), 41-47. doi:10.1007/s10334-011-0249-8.

Item is

Basisdaten

einblenden: ausblenden:
Genre: Zeitschriftenartikel

Dateien

einblenden: Dateien
ausblenden: Dateien
:
Streicher_2012_Effects.pdf (Verlagsversion), 524KB
 
Datei-Permalink:
-
Name:
Streicher_2012_Effects.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Privat
MIME-Typ / Prüfsumme:
application/pdf
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Streicher, Markus1, Autor           
Schäfer, Andreas1, Autor           
Reimer, Enrico1, Autor           
Dhital, Bibek1, Autor           
Trampel, Robert1, Autor           
Ivanov, Dimo1, Autor           
Turner, Robert1, Autor           
Affiliations:
1Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634550              

Inhalt

einblenden:
ausblenden:
Schlagwörter: Air; Susceptibility; Temperature; PRF; PRFS-based MR thermometry
 Zusammenfassung: The temperature dependence of the proton resonance frequency (PRF) is often used in MR thermometry. However, this method is prone to even very small changes in local magnetic field strength. Here, we report on the effects of susceptibility changes of surrounding air on the magnetic field inside an object and their inferred effect on the measured MR temperature. Materials and methods MR phase thermometry was performed on spherical agar phantoms enclosed in cylindrical containers at 7 T. The air susceptibility inside the cylindrical container was changed by both heating the air and changing the gas composition. Results Changing the temperature of surrounding air from 23 to 69°C caused an apparent MR temperature error of 2 K. When ambient air was displaced by 100% oxygen, the MR temperature error increased to 40 K. The magnetic field shift and therefore error in inferred MR temperature scales linearly with volume susceptibility change and has a strong and nontrivial dependence on the experimental configuration. Conclusion Air susceptibility changes associated with oxygen concentration changes greatly affect PRF MR thermometry measurements. Air temperature changes can also affect these measurements, but to a smaller degree. For uncalibrated MR thermometry, air susceptibility changes may be a significant source of error.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2011-02-112010-12-102011-03-082011-04-102012-02-01
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1007/s10334-011-0249-8
Anderer: Epub 2011
PMID: 21479876
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Magnetic Resonance Materials in Physics, Biology and Medicine
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: Amsterdam
Seiten: - Band / Heft: 25 (1) Artikelnummer: - Start- / Endseite: 41 - 47 Identifikator: ISSN: 0968-5243
CoNE: https://pure.mpg.de/cone/journals/resource/954926245532