English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT

Released

Journal Article

High-speed chiral separations on a microchip with UV detection

MPS-Authors
/persons/resource/persons58782

Ludwig,  Martin
Research Group Belder, Max-Planck-Institut für Kohlenforschung, Max Planck Society;

/persons/resource/persons58706

Kohler,  Frank
Research Group Belder, Max-Planck-Institut für Kohlenforschung, Max Planck Society;

/persons/resource/persons58418

Belder,  Detlev
Research Group Belder, Max-Planck-Institut für Kohlenforschung, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Ludwig, M., Kohler, F., & Belder, D. (2003). High-speed chiral separations on a microchip with UV detection. Electrophoresis, 24(18), 3233-3238. doi:10.1002/elps.200305427.


Cite as: https://hdl.handle.net/11858/00-001M-0000-000F-9833-2
Abstract
Fast chiral separations of a variety of basic and acidic compounds could be realized on microfluidic quartz chips. A microchip electrophoresis instrument equipped with a linear imaging UV-detector was used. The usually applied but troublesome fluorescence tagging in order to enable fluorescence detection could be omitted. Using sulfated cyclodextrins as chiral selectors baseline separation of 19 compounds could be achieved in less than 1 min with high reproducibility. The relative standard deviation of migration time was below 7%. The fastest separation could be performed in 2.5 s which is to date the fastest separation of enantiomers reported. It was possible to apply microchip electrophoresis (MCE) for the determination of high enantiomeric excess (ee) values, as exemplarily shown for pseudoephedrin where 2% of the minor enantiomer could reliably be determined beside high amount of the other isomer. Successful separation of a mixture of 3 chiral drugs could be performed in a single run in less than 11 s utilizing a separation length of only 12 mm. These results show that MCE has great potential for fast chiral analysis and high-throughput screening.